CRANI PACK, Model/Catalog Number WECR41D, medical procedure kit
American Contract Systems Inc began a recall of CRANI PACK, Model/Catalog Number WECR41D, medical procedure kit on . FDA classified it as Class II and gives the reason as "During an internal investigation, ACS was unable to confirm that the CODMAN component achieved the required Sterility Assurance Level and therefore is not suitable for sterilization according to ACS procedures". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1406-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- American Contract Systems, Inc.
- Product
- CRANI PACK, Model/Catalog Number WECR41D, medical procedure kit.
- Product code
- OJG Neurological Tray
- Reason for recall
- During an internal investigation, ACS was unable to confirm that the CODMAN component achieved the required Sterility Assurance Level and therefore is not suitable for sterilization according to ACS procedures.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Owens & Minor issued an URGENT: MEDICAL DEVICE FIELD CORRECTION notice to its consignees on 02/01/2025 via email. The notice explained the issue, risk to health, and requested the following: ACTIONS TO TAKE: " Discontinue use of the Codman Integrated Bipolar Cord and Tubing Set (Manufacturer's Item Number COD9190002RPB, ACS Item Number 347473). " Add warning labels to affected kits which contain the component to ensure the component is discarded. A labeling template is enclosed for printing labels which can be affixed on all affected kits in inventory. This label should be applied to each affected Procedure tray and product case in a prominently visible location to end -users. The location chosen should not cover any other critical product information found on existing product labeling. o Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. o Distributors: If any of the affected procedure trays identified above were further distributed, please ensure all end-users are appropriately notified of this product correction and maintain records of effectiveness. You are also responsible for the reconciliation of these customers. Please do not have your customers respond directly to ACS. " Complete the attached Recall Response Form and return as soon as possible. Report any quality problems or adverse events associated with the products listed in this notification to [email protected].
- Quantity in commerce
- 91 units
- Distribution
- US: MI
- Product codes and lots
- Lot Code: UDI: 191072216984, Lot Cocdes: 71-052536, Exp: 10/30/2025; 71-051491, Exp: 07/16/2025; LOT: 71-051609, Exp: 07/30/2025; 71-051767, Exp: 08/14/2025; 71-053065, Exp: 12/10/2025
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1406-2025 | Class II | Ongoing | Voluntary: Firm initiated |