VERICIS, Merge Cardio, Model/Catalog Number: Version 9.x, Software Version: 9.0.6 and 9.0.8; VERICIS is an...
Merge Healthcare, Inc. began a recall of VERICIS, Merge Cardio, Model/Catalog Number: Version 9.x, Software Version: 9.0.6 and 9.0.8; VERICIS is an integrated cardiovascular information system... on . FDA classified it as Class II and gives the reason as "When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1398-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Merge Healthcare Incorporated
- Product
- VERICIS, Merge Cardio, Model/Catalog Number: Version 9.x, Software Version: 9.0.6 and 9.0.8; VERICIS is an integrated cardiovascular information system classified as a picture archiving and communications system.
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Merge Healthcare issued an Urgent Field Safety Notice to it's consignees on 02/19/2025 via email. The notice explained the issue, risk to patients, provided a workaround to immediately address the issue, and recommended clinicians check measurements in the final report to ensure that they accurately reflect the patient's condition pending software update. Required Actions: Merge Healthcare Support will contact all customers to: i) Confirm that you have read and understood this Urgent Field Safety Notice. ii) Schedule an update to implement the configuration change on your system. For more information, please contact Merge Healthcare Support at 1-877-741-5369 (available 24 hours a day, 7 days a week) or visit our website: https://www.merative.com/support/merge. Please forward this notice to any individuals within your organization that may need to be aware of this notification. Customers should alert any other affiliated parties that may be affected by this Urgent Field Safety Notice. The firm identified a missing software version (11.0.2) and issued the corrected notice on 03/03/2025.
- Quantity in commerce
- 2
- Distribution
- US
- Product codes and lots
- The following versions were introduced prior to the implementation of UDI. Version Numbers: 9.0.6, 9.0.8, Part Numbers: 88-00061-00, 88-00062-00, 88-00063-00, 88-00054-00, 88-00064-00.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1398-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1399-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1400-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1401-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1402-2025 | Class II | Ongoing | Voluntary: Firm initiated |