Brand Name: Da Vinci 5 Product Name: ASSY,DV5 CONSOLE,IS5000 Foot Tray Pedals Model/Catalog Number: 380730...
Intuitive Surgical, Inc. began a recall of Brand Name: Da Vinci 5 Product Name: ASSY,DV5 CONSOLE,IS5000 Foot Tray Pedals Model/Catalog Number: 380730 Software Version: N/A . Refer to HHE for additional... on . FDA classified it as Class II and gives the reason as "Due to an increase in complaints concerning foot tray pedal spring failing resulting in the pedal remaining pressed". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1325-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Intuitive Surgical, Inc.
- Product
- Brand Name: Da Vinci 5 Product Name: ASSY,DV5 CONSOLE,IS5000 Foot Tray Pedals Model/Catalog Number: 380730 Software Version: N/A . Refer to HHE for additional details. Component: N/A
- Product code
- NAY System, Surgical, Computer Controlled Instrument
- Reason for recall
- Due to an increase in complaints concerning foot tray pedal spring failing resulting in the pedal remaining pressed
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On February 26, 2025, Intuitive Surgical issued a Urgent: Medical Device Correction to affected consignees via E-Mail. Intuitive asked consignees to take the following actions: 1. Read and understand this notification. 2. Share this notification with all personnel at your site who perform and support da Vinci procedures. 3. Inform affected personnel when the correction has been completed. 4. Complete the attached Acknowledgement Form immediately and return it via fax or email to Intuitive as instructed on the form. 5. Please retain a copy of this letter and the acknowledgement form for your files. 6. Please inform Intuitive of any Adverse Events* or quality problems concerning the use of the subject devices via the standard complaint process 7. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail or by fax. 8. Complete and submit the report Online Regular Mail or Fax: Download form or call 1- 800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form or submit by fax to 1-800-FDA-0178. An Intuitive representative will schedule a visit to replace the affected foot tray assembly.
- Quantity in commerce
- 439 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Alaska, Arizona, Arkansas, California, Colorado, Connecticut, District of Columbia, Florida, Georgia, Idaho, Illinois, Indiana, Kansas, Kentucky, Louisiana, Maryland, Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New Mexico, New York, North Carolina, North Dakota, Ohio, Oklahoma, Oregon, Pennsylvania, South Carolina, South Dakota, Tennessee, Texas, Utah, Virginia, Washington, West Virginia, Wisconsin, Wyoming and the country of South Korea.
- Product codes and lots
- UDI: 00886874119747/ All da Vinci dV5 consoles and foot trays assemblies manufactured prior to 12/20/2024.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1325-2025 | Class II | Ongoing | Voluntary: Firm initiated |