Z-1407-2025Class II

Synapse PACS Software Version 7.4.x; Software Versions: 7.4.000, 7.4.001, 7.4.010, 7.4.100, 7.4.110, 7.4.200

FUJIFILM Healthcare Americas Corporation / initiated 2025-02-11 / Open, Classified / root cause: software

FUJIFILM Healthcare Americas Corporation began a recall of Synapse PACS Software Version 7.4.x; Software Versions: 7.4.000, 7.4.001, 7.4.010, 7.4.100, 7.4.110, 7.4.200 on . FDA classified it as Class II and gives the reason as "The incorrect computed patient age is showing in VX for patients less than 3 months old". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-1407-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
FUJIFILM Healthcare Americas Corporation
Product
Synapse PACS Software Version 7.4.x; Software Versions: 7.4.000, 7.4.001, 7.4.010, 7.4.100, 7.4.110, 7.4.200.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
The incorrect computed patient age is showing in VX for patients less than 3 months old.
Root cause tag
software
FDA root cause
Software change control
Action
On February 11, 2025 URGENT: MEDICAL DEVICE RECALL letters were sent to customers. ACTIONS TO BE TAKEN BY THE COMPANY Fujifilm will issue corrections for all impacted 7.4.x versions. The following patch releases will be made available: - 7.4.020 for 7.4.0x versions (immediately available) - 7.4.120 for 7.4.1x versions (available February 28, 2025) - 7.4.210 for 7.4.2x versions (available February 28,2025) ACTIONS THAT CAN BE TAKEN BY THE END-USER IMMEDIATELY Regarding this issue, Synapse PACS VX can safely be used by the end-user as long as; 1] the end-user is aware of the issue and, 2] takes appropriate actions when rendering a diagnosis for patients under 18 years of age. The following workarounds can be used: 1) The image acquisition date and date of birth are displayed on the Image. The end-user can use the dates to verify the age or, 2) Open the Series Viewer by double clicking on the thumbnail displayed on top the Synapse VX Image viewing window or the worklist. The Series Viewer Window will render the image on the server using the Server-side rendering (Zero Viewer component) which will calculate the age correctly. Alternatively, the user can switch to the Synapse PACS Zero Viewer/Thin Viewer (Server-side Rendering). The Synapse PACS Zero Viewer/Thin Viewer calculates the age correctly. ACTIONS TO BE TAKEN BY THE CUSTOMER/END-USER 1) Please complete and return the Field Action Verification Form attached to this communication. Responding with the information requested is essential for ensuring appropriate action is taken. 2) Please contact Customer Service at 1-888-FUJIMED (1-888-385-4633) to schedule a patch update. Fujifilm regrets any inconvenience this action may cause and appreciates your understanding and cooperation. Please contact our field action coordinator, Jay Wigley, via email at [email protected], if you have any further questions regarding this field action. Please contact Technical Assistance Center (TAC) at 1-888-385-4622 or 1-888-FU
Quantity in commerce
104 units (69 US, 35 OUS)
Distribution
Domestic: AR, AZ, CA, CO, CT, FL, GA, HI, IL, IN, KS, KY, LA, MA, MN, MO, MS, NC, NJ, NY, OH, OK, OR, TN, TX, VT, WA, WI, WV, WY. International: Australia, Belgium, Brazil, France, Germany, Portugal, UK
Product codes and lots
Model numbers: Synapse PACS 7.4.000, 7.4.001, 7.4.010, 7.4.100, 7.4.110, 7.4.200 Software Version Numbers: Synapse PACS 7.4.000, UDI: (01)00854904006008(10)0704000; Synapse PACS 7.4.001, UDI: (01)00854904006008(10)0704001; Synapse PACS 7.4.010, UDI: (01)00854904006008(10)0704010; Synapse PACS 7.4.100, UDI: (01)00854904006008(10)0704100; Synapse PACS 7.4.110, UDI: (01)00854904006008(10)0704110; Synapse PACS 7.4.200, UDI: (01)00854904006008(10)0704200.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1407-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1407-2025Class IIOngoingVoluntary: Firm initiated