Medline procedure kits labeled as: 1) ARTHROSCOPY PACK, Pack Number DYNJHS0101I; 2) ARTHROSCOPY PACK-LF, Pack...
MEDLINE INDUSTRIES, LP - Northfield began a recall of Medline procedure kits labeled as: 1) ARTHROSCOPY PACK, Pack Number DYNJHS0101I; 2) ARTHROSCOPY PACK-LF, Pack Number DYNJ0390808F; 3) ARTHROSCOPY PACK-LF, Pack... on . FDA classified it as Class II and gives the reason as "Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/other quality issues have been identified that may pose a risk to patient health". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-1452-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Product
- Medline procedure kits labeled as: 1) ARTHROSCOPY PACK, Pack Number DYNJHS0101I; 2) ARTHROSCOPY PACK-LF, Pack Number DYNJ0390808F; 3) ARTHROSCOPY PACK-LF, Pack Number DYNJ23150J; 4) ARTHROSCOPY PACK-LF, Pack Number DYNJ21233O; 5) ARTHROSCOPY PACK-LF, Pack Number DYNJ21233O; 6) ARTHROSCOPY TRAY, Pack Number DYNJ46435A; 7) KNEE ARTHROSCOPY PACK, Pack Number DYNJ50120D
- Product code
- OII Arthrogram Tray
- Reason for recall
- Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/other quality issues have been identified that may pose a risk to patient health.
- FDA root cause
- Under Investigation by firm
- Action
- Medline Industries LP issued a recall notice to its consignees on 02/11/2025 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request the syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other problems, and report any quality issues to Medline and the FDA. Distributors were directed to notify their customers. This is an expansion of RES 94583
- Quantity in commerce
- 186 units
- Distribution
- Worldwide distribution - US Nationwide and the country of Canada.
- Product codes and lots
- 1) Pack Number DYNJHS0101I: UDI/DI 10888277046412 (each) 40888277046413 (case), Lot Number 21LBA780A; 2) Pack Number DYNJ0390808F: UDI/DI 10889942244591 (each) 40889942244592 (case), Lot Number 21JMG164A; 3) Pack Number DYNJ23150J: UDI/DI 10888277742239 (each) 40888277742230 (case), Lot Number 21LMF991A; 4) Pack Number DYNJ21233O: UDI/DI 10889942543878 (each) 40889942543879 (case), Lot Number 22GBQ106M; 5) Pack Number DYNJ21233O: UDI/DI 10889942543878 (each) 40889942543879 (case), Lot Number 22JBA123Z; 6) Pack Number DYNJ46435A: UDI/DI 10889942670901 (each) 40889942670902 (case), Lot Number 23LMG014A; 7) Pack Number DYNJ50120D: UDI/DI 10195327195656 (each) 40195327195657 (case), Lot Number 22JBG571Z
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1435-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1436-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1437-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1438-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1439-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1440-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1441-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1442-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1443-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1444-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1445-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1446-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1447-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1448-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1449-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1450-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1451-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1452-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1453-2025 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.