Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1408-2025Class II

Premier Solo Diamond - Large Invented Cone; SKU: 807016C

Premier Dental Products Co / initiated 2025-02-27 / Open, Classified / root cause: manufacturing process

Premier Dental Products Co began a recall of Premier Solo Diamond - Large Invented Cone; SKU: 807016C on . FDA classified it as Class II and gives the reason as "The hardness not meeting the material specification and may cause the diamond bur to bend". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1408-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Premier Dental Products Co
Product
Premier Solo Diamond - Large Invented Cone; SKU: 807016C.
Product code
EJL Bur, Dental
Reason for recall
The hardness not meeting the material specification and may cause the diamond bur to bend.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On February 25, 2025 MEDICAL DEVICE RECALL letters were sent to customers. Actions to be taken: Immediately examine your inventory and quarantine all product subject to this recall, including individual bottles in partial boxes. This recall should be carried out to the healthcare practitioner level. If you have further distributed this product, please identify and notify all your customers that received the affected SKU and lot numbers of this product at once. You may include a copy of this notification with your communication to your customers that have received affected product. Please have your customers account for all product in their possession and arrange for the return of all affected product. Your customers should return the affected product to you for consolidation first prior to returning to Premier Dental Products Company. Please complete the enclosed Required Recall Notice Response form and return to Premier immediately via email ([email protected]) or mail with the returned product to: Premier Solo Diamond Attn: Jessica Huang Premier Dental Products Company 1710 Romano Drive Plymouth Meeting, PA 19462 Upon receipt of the returned product at Premier Dental, Premier s Customer Experience can issue a credit or Premier can replace the returned item with SKU 807018C - SOLO Diamond Large Inverted Cone. If you have any further questions regarding billing or replacements, please contact Customer Experience at [email protected]. We thank you for your assistance in completing this activity and apologize for any inconvenience that this may cause.
Quantity in commerce
77 units
Distribution
US Nationwide distribution in the states of CA, MN, NY, PA, TN, VA, WA.
Product codes and lots
SKU: 807016C; UDI-DI: 00348783000237; Lot Numbers: M0117646, M0120260, M0123294;

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1408-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1408-2025Class IIOngoingVoluntary: Firm initiated