DxI 9000 Access Immunoassay Analyzer, catalog number C11137 in vitro diagnostic device used for the...

Siemens Healthcare Diagnostics Inc. / initiated 2024-07-12 / Open, Classified / root cause: design

Beckman Coulter, Inc. began a recall of DxI 9000 Access Immunoassay Analyzer, catalog number C11137 in vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative... on . FDA classified it as Class II and gives the reason as "Ground screws on Dxl 9000 PCB boards may cause electric shorts, leading to smoke and potentially delaying patient results. Smoke also poses a hazard to equipment and users". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-1529-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
DxI 9000 Access Immunoassay Analyzer, catalog number C11137 in vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in human body fluids
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Ground screws on Dxl 9000 PCB boards may cause electric shorts, leading to smoke and potentially delaying patient results. Smoke also poses a hazard to equipment and users.
Root cause tag
design
FDA root cause
Device Design
Action
The firm sent notice of "Urgent Medical Device Recall" to affected consignees (all OUS), dated July 3, 2024. Customers were informed of the potential for barcode read failure error, delay of reporting patient test results, and possibility for a small amount of smoke to be emitted. Consignees were advised that no action was necessary if their laboratory has not observed any RSM Barcode Read Failure error messages. If you do observe any such error messages, or notice smoking or burning smell coming from the Dxl 9000 analyzer, please contact your Beckman Coulter service representative. Beckman Coulter will no longer ship these analyzers with the affected ground screw. Your Beckman Coulter service representative will contact you to schedule a part replacement that resolves the issue.
Quantity in commerce
30 OUS
Distribution
OUS (Foreign) distribution to countries of: Australia, Croatia, Czechia, France, Germany, Ireland, Israel, Italy, Netherlands, New Zealand, South Africa, Spain, and United Kingdom of Great Britain and Northern Ireland.
Product codes and lots
UDI-DI 15099590732103 Serial/Lot Numbers: 300116, 300117, 300118, 300123, 300126, 300128, 300129, 300130, 300131, 300133, 300134, 300135, 300136, 300137, 300138, 300139, 300141, 300142, 300143, 300145, 300147, 300148, 300149, 300150, 300151, 300152, 300153, 300154, 300155, and 300156

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1529-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1529-2025Class IIOngoingVoluntary: Firm initiated