DxI 9000 Access Immunoassay Analyzer, catalog number C11137 in vitro diagnostic device used for the...
Beckman Coulter, Inc. began a recall of DxI 9000 Access Immunoassay Analyzer, catalog number C11137 in vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative... on . FDA classified it as Class II and gives the reason as "Ground screws on Dxl 9000 PCB boards may cause electric shorts, leading to smoke and potentially delaying patient results. Smoke also poses a hazard to equipment and users". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1529-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- DxI 9000 Access Immunoassay Analyzer, catalog number C11137 in vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in human body fluids
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- Ground screws on Dxl 9000 PCB boards may cause electric shorts, leading to smoke and potentially delaying patient results. Smoke also poses a hazard to equipment and users.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The firm sent notice of "Urgent Medical Device Recall" to affected consignees (all OUS), dated July 3, 2024. Customers were informed of the potential for barcode read failure error, delay of reporting patient test results, and possibility for a small amount of smoke to be emitted. Consignees were advised that no action was necessary if their laboratory has not observed any RSM Barcode Read Failure error messages. If you do observe any such error messages, or notice smoking or burning smell coming from the Dxl 9000 analyzer, please contact your Beckman Coulter service representative. Beckman Coulter will no longer ship these analyzers with the affected ground screw. Your Beckman Coulter service representative will contact you to schedule a part replacement that resolves the issue.
- Quantity in commerce
- 30 OUS
- Distribution
- OUS (Foreign) distribution to countries of: Australia, Croatia, Czechia, France, Germany, Ireland, Israel, Italy, Netherlands, New Zealand, South Africa, Spain, and United Kingdom of Great Britain and Northern Ireland.
- Product codes and lots
- UDI-DI 15099590732103 Serial/Lot Numbers: 300116, 300117, 300118, 300123, 300126, 300128, 300129, 300130, 300131, 300133, 300134, 300135, 300136, 300137, 300138, 300139, 300141, 300142, 300143, 300145, 300147, 300148, 300149, 300150, 300151, 300152, 300153, 300154, 300155, and 300156
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1529-2025 | Class II | Ongoing | Voluntary: Firm initiated |