InPen smart insulin pen, REF: MMT-105NNBLNA, MMT-105NNGYNA, MMT-105NNPKNA, MMT-105ELBLNA, MMT-105ELGYNA...
Medtronic MiniMed, Inc. began a recall of InPen smart insulin pen, REF: MMT-105NNBLNA, MMT-105NNGYNA, MMT-105NNPKNA, MMT-105ELBLNA, MMT-105ELGYNA, MMT105ELPKNA on . FDA classified it as Class II and gives the reason as "Insulin pens may have been incorrectly assembled therefore users could experience the following: either the insulin cartridge will not fit into the cartridge holder; or the cartridge holder may be difficult to remove...". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-1650-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Medtronic, Inc.
- Product
- InPen smart insulin pen, REF: MMT-105NNBLNA, MMT-105NNGYNA, MMT-105NNPKNA, MMT-105ELBLNA, MMT-105ELGYNA, MMT105ELPKNA
- Product code
- FMF Syringe, Piston
- Reason for recall
- Insulin pens may have been incorrectly assembled therefore users could experience the following: either the insulin cartridge will not fit into the cartridge holder; or the cartridge holder may be difficult to remove from the insulin pen. If the issue is present, and if the user does not have backup insulin they may experience temporary hyperglycemia.
- FDA root cause
- Mixed-up of materials/components
- Action
- On 3/4/2025, correction notices were sent to customers and distributors informing them of the following: 1) If you have already inserted your insulin cartridge into your device without difficulty removing the cartridge holder, then you can continue normal use. 2) If you have a device from one of the affected lot numbers and have not yet started using it, or have had difficulties in removing the cartridge holder or inserting the insulin cartridge, please call 1-800-646-4633, option 1, or go to https://info.medtronicdiabetes.com/ICH-replacement-form to request a replacement product. Do not use this device and dispose of it according to local regulations. 3) Complete and return the confirmation form via link: https://info.medtronicdiabetes.com/cartridge-holder Distributors are asked to notify impacted customers. If you have further questions or need assistance, please call the firm's 24-Hour Technical Support line at 1-800-646-4633, option 1.
- Quantity in commerce
- 8,986
- Distribution
- US: OK, IA, NC, SC, MI, MN, CA, AZ, AL, MO, WA, IL, KS, TX, FL, MT, MD, OH, VA, NY, TN, HI, MA, NJ, PA, CO, WI, GA, CT, NE, OR, LA, NM, SD, ND, IN, NV, WV, RI, AR, NH
- Product codes and lots
- REF/UDI-DI(GTIN)/Lot: MMT-105NNBLNA/763000827182/D0020, D0034, D0036, D0038, D0045, D0046, D0050, D0057; MMT-105NNGYNA/763000827199/D0026, D0037, D0039, D0041, D0042, D0043, D0049; MMT-105NNPKNA/763000827205/D0029, D0030, D0040, D0051, D0062; MMT-105ELBLNA/763000827151/D0021, D0033, D0035, D0044, D0059, D0060, D0061; MMT-105ELGYNA/763000827168/D0031, D0032, D0047, D0068; MMT105ELPKNA/763000827175/D0027, D0028, D0048
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1650-2025 | Class II | Ongoing | Voluntary: Firm initiated |