Z-1324-2025Class II
Spectral CT on Rails, Software Version Number 5.1.0, Model number: 728334
Philips Medical Systems Nederland B.V. / initiated 2025-02-28 / Open, Classified / root cause: software
Philips Medical Systems Nederland B.V. began a recall of Spectral CT on Rails, Software Version Number 5.1.0, Model number: 728334 on . FDA classified it as Class II and gives the reason as "Multiple problems identified with the software version leading to various scanning and image issues, and unintended device movement". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1324-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Philips Medical Systems Nederland B.V.
- Product
- Spectral CT on Rails, Software Version Number 5.1.0, Model number: 728334;
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- Multiple problems identified with the software version leading to various scanning and image issues, and unintended device movement.
- Root cause tag
- software
- FDA root cause
- Under Investigation by firm
- Action
- On February 28, 2025 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Actions that should be taken by the customer / user in order to prevent risks for patients or users. " You may continue to use your system(s) in accordance with the intended use and by following the recommendation listed below. " When you see the warning message No communication with table and C-arm. Avoid collisions, Proceed with caution., acknowledge the message by selecting OK and continue using the CT while observing with extra care, collision with the Gantry and C-arm can be avoided. " To protect the patient and the operator from a potential collision due to an automatic gantry motion during scan and plan box not updated during interventional procedure issues, follow the instructions mentioned in the IFU Section 4 Preparing for an Examination and Sub section Gantry Operation: o During all movements of the gantry (automatic and manual), keep the patient under continuous observation. o Take care to avoid pressing the patient against the gantry or between table parts, as well as to avoid disconnecting any infusion or resuscitation apparatus. o For all patient table types, take care when using attachments to avoid collision with the gantry. o Make sure that the patient is strapped securely to avoid dangling of the hands. Ensure that the patient is placed securely on the patient table and is not in danger of falling. o Auto scan means that automatic motions are expected without using the enable button. Ensure the patient table does not collide with the gantry during movement. o During studies, the gantry movements (if applicable) are automatic. Ensure enough clearance between the patient and the gantry. Before initiating the scan, perform manual movements to check the clearance. " To avoid cone beam auto correction artifacts: 1) Reduce the collimation to 32 for brain study, or reduce the collimation to 112 (or any value below 128) for body imaging, which will turn CBAC off. Refer t
- Quantity in commerce
- 1 unit
- Distribution
- Domestic: MN
- Product codes and lots
- Software Version Number 5.1.0; Model number: 728334; UDI-DI: (01)00884838103627(21)1005; Serial Number (SN): 1005;
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1324-2025 | Class II | Ongoing | Voluntary: Firm initiated |