Welch Allyn CP150 Electrocardiograph
Baxter Healthcare Corporation began a recall of Welch Allyn CP150 Electrocardiograph on . FDA classified it as Class II and gives the reason as "There were customer reports of devices which experienced battery-related fires. The investigation found that the devices in question were utilizing third-party, after-market batteries that are not approved by Baxter". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1560-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Baxter Healthcare Corporation
- Product
- Welch Allyn CP150 Electrocardiograph:
- Product code
- DPS Electrocardiograph
- Reason for recall
- There were customer reports of devices which experienced battery-related fires. The investigation found that the devices in question were utilizing third-party, after-market batteries that are not approved by Baxter.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- An URGENT MEDICAL DEVICE CORRECTION notification letter dated 3/24/25 was sent to customers. Actions to be Taken by Customers 1. DO NOT USE UNAPPROVED BATTERIES. Baxter asks customers to only use the approved Baxter / Welch Allyn BATT99 battery, consistent with product labeling due to their safety and performance testing. 2. INSPECT ALL DEVICES. If you have a device identified in the affected product table above, check the device for the correct battery type. See Figure 1 below for examples of labels from approved batteries labelled with the BATT99 product code, Welch Allyn , and the Skaneateles Falls manufacturing facility address. 3. STOP USING THE DEVICE if an unapproved battery is found. 4. IMMEDIATELY REPLACE THE UNAPPROVED BATTERY with an approved battery. Please contact Baxter to order the approved Baxter / Welch Allyn BATT99 batteries. Batteries can be ordered at the following link: https://parts.hillrom.com/hillromUS/en 5. Please forward a copy of this communication to the Medical Director/Office Manager, Director of Biomedical Engineering, Patient Safety Officer, Risk Manager, Surgical Director and any other departments within your institution who use the affected product. 6. If you received this communication directly from Baxter, acknowledge receipt by following the instructions on the enclosed reply instruction sheet, even if you have no remaining inventory. Acknowledging receipt of this notification will prevent you from receiving repeat notices. If you do not complete the acknowledgement, you will receive a phone call from OnProcess Technology on behalf of Baxter to confirm your receipt of this notification. 7. If you purchased this product from a distributor or wholesaler, please note that responding on the Baxter customer portal is not applicable. If a response is requested by your distributor or wholesaler, please respond to them according to their instructions. 8. If you are a dealer/wholesaler, distributor/reseller or Original Equipment Manufac
- Quantity in commerce
- 41,739 units
- Distribution
- Worldwide - US Nationwide distribution.
- Product codes and lots
- All Serial Numbers. Product Code: CP150-2ES2, UDI-DI: 00732094192131; Product Code: CP150-2FR2, UDI-DI: 00732094192124; Product Code: CP150-2FR5, UDI-DI: 00732094192117; Product Code: CP150-2NL2, UDI-DI: 00732094192100; Product Code: CP150-2PT2, UDI-DI: 00732094192087; Product Code: CP150-2SV2, UDI-DI: 00732094192070; Product Code: CP150-3ENB, UDI-DI: 00732094179705; Product Code: CP150-3ESB, UDI-DI: 00732094179682; Product Code: CP150-4DE2, UDI-DI: 00732094192063; Product Code: CP150-4EN2, UDI-DI: 00732094179675; Product Code: CP150-4EN4, UDI-DI: 00732094179668; Product Code: CP150-4PT2, UDI-DI: 00732094192025; Product Code: CP150A-1EN2, UDI-DI: 00732094185157; Product Code: CP150A-1EN4, UDI-DI: 00732094185140; Product Code: CP150A-1EN66, UDI-DI: 00732094179644; Product Code: CP150A-1ENB, UDI-DI: 00732094177633; Product Code: CP150A-1ENBR, UDI-DI: 00732094322415; Product Code: CP150A-1ESB, UDI-DI: 00732094192230; Product Code: CP150A-1FRB, UDI-DI: 00732094179637; Product Code: CP150A-2DE2, UDI-DI: 00732094192018; Product Code: CP150A-2DE5, UDI-DI: 00732094192001; Product Code: CP150A-2EN2, UDI-DI: 00732094177626; Product Code: CP150A-2EN3, UDI-DI: 00732094179620; Product Code: CP150A-2EN4, UDI-DI: 00732094177619; Product Code: CP150A-2EN7, UDI-DI: 00732094191998; Product Code: CP150A-2ES2, UDI-DI: 00732094191981; Product Code: CP150A-2FR2, UDI-DI: 00732094191974; Product Code: CP150A-2FR5, UDI-DI: 00732094191967; Product Code: CP150A-2NL2, UDI-DI: 00732094191950; Product Code: CP150A-2NO2, UDI-DI: 00732094191943; Product Code: CP150A-2SV2, UDI-DI: 00732094191929; Product Code: CP150A-3BPZ, UDI-DI: 00732094212938; Product Code: CP150A-3EN2, UDI-DI: 00732094179613; Product Code: CP150A-3EN3, UDI-DI: 00732094179606; Product Code: CP150A-3EN4, UDI-DI: 00732094184815; Product Code: CP150A-3ENB, UDI-DI: 00732094179590; Product Code: CP150A-3ES2, UDI-DI: 00732094179583; Product Code: CP150A-3ESB, UDI-DI: 00732094177602; Product Code: CP150A-3ZHC-CN, UDI-DI: 00732094213515; Product Code: CP150A-4BPZ, UDI-DI: 00732094212952; Product Code: CP150A-4DE2, UDI-DI: 00732094191912; Product Code: CP150A-4EN2, UDI-DI: 00732094177596; Product Code: CP150A-4EN4, UDI-DI: 00732094177589; Product Code: CP150A-4ES2, UDI-DI: 00732094177572; Product Code: CP150A-4FR5, UDI-DI: 00732094189186; Product Code: CP150A-4PT2, UDI-DI: 00732094189179; Product Code: CP150A-4ZHC-CN, UDI-DI: 00732094213539; Product Code: CP150AD-1ENB, UDI-DI: 00732094237580; Product Code: CP150AD-2DE2, UDI-DI: 00732094237573; Product Code: CP150AD-2ES2, UDI-DI: 00732094237559; Product Code: CP150ADS-1EN66, UDI-DI: 00732094242171; Product Code: CP150AS-1EN66, UDI-DI: 00732094333244; Product Code: CP150AS-1ENB, UDI-DI: 00732094185133; Product Code: CP150AW-1EN66, UDI-DI: 00732094237511; Product Code: CP150-2EN7, UDI-DI: 00732094192148; Product Code: CP150-1EN2, UDI-DI: 00732094185171; Product Code: CP150-1EN66, UDI-DI: 00732094179750; Product Code: CP150-1ENB, UDI-DI: 00732094177640; Product Code: CP150-1FRB, UDI-DI: 00732094192254; Product Code: CP150-2DE2, UDI-DI: 00732094192162; Product Code: CP150-2EN2, UDI-DI: 00732094179743; Product Code: CP150-2EN4, UDI-DI: 00732094179729; Product Code: CP150AW-1ENB, UDI-DI: 00732094237504; Product Code: CP150AW-2DE2, UDI-DI: 00732094237498; Product Code: CP150AW-2EN4, UDI-DI: 00732094237481; Product Code: CP150AW-2ES2, UDI-DI: 00732094237474; Product Code: CP150AW-2NL2, UDI-DI: 00732094237467; Product Code: CP150AW-3ENB, UDI-DI: 00732094242003; Product Code: CP150AWD-1EN66, UDI-DI: 00732094237436; Product Code: CP150AWD-1ENB, UDI-DI: 00732094237429; Product Code: CP150AWD-2DE2, UDI-DI: 00732094237412; Product Code: CP150AWD-2EN4, UDI-DI: 00732094237405; Product Code: CP150AWD-2ES2, UDI-DI: 00732094237399; Product Code: CP150AWD-2NL2, UDI-DI: 00732094237382; Product Code: CP150AWD-4EN4, UDI-DI: 00732094237368; Product Code: CP150AWDS-1ENB, UDI-DI: 00732094241686; Product Code: CP150AWDS-3ENB, UDI-DI: 00732094241662; Product Code: CP150AWS-1ENB, UDI-DI: 00732094243437; Product Code: CP150S-1ENB, UDI-DI: 00732094222302; Product Code: CP150W-1ENB, UDI-DI: 00732094232837;
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1558-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1559-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1560-2025 | Class II | Ongoing | Voluntary: Firm initiated |