Brand Name: ID NOW RSV 24T Product Name: ID NOW RSV 24T Model/Catalog Number: 435-000 Software Version: N/A...
Abbott Diagnostics Scarborough, Inc. began a recall of Brand Name: ID NOW RSV 24T Product Name: ID NOW RSV 24T Model/Catalog Number: 435-000 Software Version: N/A Product Description: Respiratory virus panel... on . FDA classified it as Class II and gives the reason as "The impacted lots have a higher occurrence of invalid rates when compared to the product Instructions for Use". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1573-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Abbott Diagnostics Scarborough, Inc.
- Product
- Brand Name: ID NOW RSV 24T Product Name: ID NOW RSV 24T Model/Catalog Number: 435-000 Software Version: N/A Product Description: Respiratory virus panel nucleic acid assay system Component: No
- Product code
- OCC Respiratory Virus Panel Nucleic Acid Assay System
- Reason for recall
- The impacted lots have a higher occurrence of invalid rates when compared to the product Instructions for Use.
- Root cause tag
- labeling and UDI
- FDA root cause
- Process change control
- Action
- On March 4, 2025 FIELD CORRECTION NOTICE letters were sent to end user customers. Please complete the following actions, as applicable. 1. If you have impacted inventory in stock, then: " Discontinue use of and destroy any remaining inventory of the impacted lots according to your procedures. " Complete and return the Customer Reply Form (Form must be completed, signed and returned to receive a replacement lot). " Please retain this letter for your records. 2. If you have forwarded the product listed above to others in your network, then: - Inform them of this Field Correction Notice, provide to them a copy of this notice and request they take the necessary action. 3. If you do not have impacted inventory in stock, then: - All product lots not identified in Appendix I can continue to be used. If you have questions regarding this information, please contact Sedgwick by phone at 888 671 8858 (available 8am 5pm EST, Monday - Friday). Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online (http://www.fda.gov/MedWatch/report.htm), by mail (http://www.fda.gov/MedWatch/getforms.htm), by phone (1-800-332-1088), or by fax (1-800-FDA-0178). If you have experienced any patient or user injury associated with this Field Correction Notice, please immediately report the event to your local area Customer Service. ****Update added 5/8/2025**** On April 29, 2025 FIELD CORRECTION NOTICE letters were sent to customers informing them of additional lots included in this recall.
- Quantity in commerce
- 178056 eaches (147472 US, 31584 OUS)
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia; Austria; Belgium; Bulgaria; Canada; Costa Rica; France; French Polynesia; Germany; Italy; Luxembourg; Malaysia; Netherlands; Paraguay; Poland; Qatar; Saudi Arabia; Slovenia; South Korea; Spain; Switzerland; Taiwan; United Kingdom; Uruguay.
- Product codes and lots
- Model 435-000; GTIN/DI 10811877010521; Lot 000M906790 Model 435-000; GTIN/DI 10811877010521; Lot 000M907271 Model 435-000; GTIN/DI 10811877010521; Lot 000M923531 Model 435-000; GTIN/DI 10811877010521; Lot 000M923879 Model 435-000; GTIN/DI 10811877010521; Lot 000M924414 Model 435-000; GTIN/DI 10811877010521; Lot 000M924430 Model 435-000; GTIN/DI 10811877010521; Lot 000M926642 Model 435-000; GTIN/DI 10811877010521; Lot 000M926679 Model 435-000; GTIN/DI 10811877010521; Lot 000M927161
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1571-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1572-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1573-2025 | Class II | Ongoing | Voluntary: Firm initiated |