Z-1578-2025Class II

Brand Name: Rover Product Name: Mobile X-ray System Model/Catalog Number: MXU-RV71 Product Description: Rover...

Micro-X Ltd. / initiated 2025-03-28 / Open, Classified / root cause: design

Micro-X Ltd. began a recall of Brand Name: Rover Product Name: Mobile X-ray System Model/Catalog Number: MXU-RV71 Product Description: Rover Mobile X-rays System ("Rover") is a mobile... on . FDA classified it as Class II and gives the reason as "Potential for early life x-ray tube failure for mobile x-ray system due to excessive thermal load on the anode". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-1578-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Micro-X Ltd.
Product
Brand Name: Rover Product Name: Mobile X-ray System Model/Catalog Number: MXU-RV71 Product Description: Rover Mobile X-rays System ("Rover") is a mobile imaging system that incorporates a self-contained X-ray generator, imaging display and software for acquiring medical diagnostic images outside a standard X-ray room. It is designed specifically for Digital Radiography (DR) with vendor-supported detectors. The Rover is designed to perform radiographic X-ray examinations for disease/injury detection, diagnosis, medical and surgical treatment planning, and therapy monitoring, where patient condition contraindicates transport to a fixed X-ray unit The Rover can be used on pediatric and adult patients, in all patient treatment areas Component: N/A
Product code
IZL System, X-Ray, Mobile
Reason for recall
Potential for early life x-ray tube failure for mobile x-ray system due to excessive thermal load on the anode.
Root cause tag
design
FDA root cause
Device Design
Action
On March 28, 2025, Micro-X Ltd. issued a "Urgent: Medical Device Correction" notification to affected consignees via E-Mail. Micro-X asked consignees to take the following actions: 1. Continue to use your product as normal in accordance with its intended use. 2. Your local service representative will be in contact to apply this update to the systems at the next site visit. 3. Circulate this notice to all users of this device so they are aware of the issue. 4. If the product has been further distributed to other facilities, please ensure this notice is immediately shared with them. 5. Please retain this letter with your system(s) documentation. 6. Please complete and return the attached response form to Micro-X to the Micro-X LTD in Australia: [email protected] and copy [email protected].
Quantity in commerce
31 units
Distribution
US Nationwide distribution in the states of CA, NJ, and Puerto Rico (US territory).
Product codes and lots
Serial Number/UDI: MSN0334 (01)09357123000051(11)220921(21)00334; MSN0350 (01)09357123000037(11)230328(21)00350 ; MSN0352 (01)09357123000051(11)230823(21)00352; MSN0364 01)09357123000051(11)230705(21)00364; MSN0382 (01)09357123000051(11)230809(21)00382; MSN0383 (01)09357123000051(11)230814(21)00383; MSN0384 (01)09357123000051(11)230816(21)00384; MSN0385 (01)09357123000051(11)230906(21)00385; MSN0386 (01)09357123000051(11)230908(21)00386; MSN0387 (01)09357123000051(11)230914(21)00387; MSN0388 (01)09357123000051(11)230916(21)00388; MSN0389 (01)09357123000051(11)230918(21)00389; MSN0390 (01)09357123000051(11)231013(21)00390; MSN0395 (01)09357123000051(11)231106(21)00395; MSN0397 (01)09357123000051(11)231118(21)00397; MSN0399 (01)09357123000051(11)240122(21)00399; MSN0400 (01)09357123000051(11)240207(21)00400; MSN0401 (01)09357123000051(11)240131(21)00401; MSN0402 01)09357123000051(11)240202(21)00402; MSN0403 (01)09357123000051(11)240205(21)00403; MSN0405 (01)09357123000051(11)240206(21)00405; MSN0408 (01)09357123000051(11)240308(21)00408; MSN0409 (01)09357123000051(11)240312(21)00409; MSN0410 (01)09357123000051(11)240318(21)00410; MSN0411 (01)09357123000051(11)240319(21)00411; MSN0422 (01)09357123000051(11)240724(21)00422; MSN0423 (01)09357123000051(11)240808(21)00423; MSN0424 (01)09357123000051(11)240809(21)00424; MSN0425 (01)09357123000051(11)240812(21)00425; MSN0426 (01)09357123000051(11)240812(21)00426; MSN0427 (01)09357123000051(11)240812(21)00427;

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1578-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1578-2025Class IIOngoingVoluntary: Firm initiated
Z-1579-2025Class IIOngoingVoluntary: Firm initiated