epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515...
Siemens Healthcare Diagnostics Inc began a recall of epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515; epoc Software Version & Sensor Configuration: epoc... on . FDA classified it as Class II and gives the reason as "The occurrence of discrepant high pH results in samples introduced with higher injection volumes with epoc sensor configuration 45.n". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1684-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515; epoc Software Version & Sensor Configuration: epoc Host SW v3.41.2, epoc NXS SW v4.14.9 & 4.14.11, Sensor Configuration 45.n.
- Product code
- CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph
- Reason for recall
- The occurrence of discrepant high pH results in samples introduced with higher injection volumes with epoc sensor configuration 45.n.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Consignees were sent an Urgent Medical Device Correction notice dated March 2025. The notification instructs consignees to review the provided notice with their medical director to determine the appropriate course of action. If affected devices were further distributed consignees are asked to forward the recall notification. Consignees are to return the provided Field Correction Effectiveness Check form within 7 days of receipt. The firm recommends that consignees stop injecting the sample as soon as the audio/visual cues are presented to minimize the potential for erroneous pH results. The current affected software expires on 06/10/25 and the upgrade will include a correction for this issue. Consignees are to retain the recall notification with their manual or near the system for reference. Consignees with questions can call the Siemens Healthineers Technical Solutions Center at 877-229-3711.
- Quantity in commerce
- 264,252 units
- Distribution
- Domestic: Nationwide Distribution; Foreign: Albania, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia, Brazil, Cambodia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Ecuador, Estonia, France, Germany, Ghana, Greece, Guadeloupe, Guatemala, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kuwait, Lithuania, Malaysia, Mali, Martinique, Mexico, Morocco, Netherlands, New Caledonia, New Zealand, Nigeria, Oman, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russian Federation, San Marino, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Vietnam.
- Product codes and lots
- Siemens Material Number (SMN): 10736515; Udi-DI: 00809708121860; Lot Numbers: All lot numbers with prefix (xx) 02 or 12; epoc Software Version and Senor Configuration: epoc Host SW V3.41.2, epoc NXS SW v4.14.9 & 4.14.11, Sensor Configuration 45.n.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1684-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1685-2025 | Class II | Ongoing | Voluntary: Firm initiated |