Z-1545-2025Class II

BD Pyxis" MedBank" Mini and BD Pyxis" MedBank" Tower, BD Pyxis" MedFlex, BD Pyxis" MedBank" MedPass are...

CareFusion 303, Inc / initiated 2025-03-11 / Open, Classified / root cause: software

CareFusion 303, Inc. began a recall of BD Pyxis" MedBank" Mini and BD Pyxis" MedBank" Tower, BD Pyxis" MedFlex, BD Pyxis" MedBank" MedPass are automated dispensing cabinets (ADC) intended to... on . FDA classified it as Class II and gives the reason as "Due to a software bug, automated dispensing cabinet is unable to calculate countback discrepancies during the issue process, which causes incorrect transaction quantity to be recorded. This bug can also result in the...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-1545-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
CareFusion 303, Inc
Product
BD Pyxis" MedBank" Mini and BD Pyxis" MedBank" Tower, BD Pyxis" MedFlex, BD Pyxis" MedBank" MedPass are automated dispensing cabinets (ADC) intended to securely store and dispense medications to a qualified and authorized user in a clinical settings.
Product code
BRY Cabinet, Table And Tray, Anesthesia
Reason for recall
Due to a software bug, automated dispensing cabinet is unable to calculate countback discrepancies during the issue process, which causes incorrect transaction quantity to be recorded. This bug can also result in the CUBIE opening more times than needed for the transaction to be completed, which may result in a patient receiving more than the intended dose of medication.
Root cause tag
software
FDA root cause
Software design
Action
On 3/11/2025, "URGENT MEDICAL DEVICE CORRECTION" notices were mailed and emailed to Medical Director, Risk Manager, Medical Device Safety Officer, Lab Manager, Clinical Engineering, and other customers asking them to take the following actions: 1) Best practice is to verify the total dose being removed is accurate against the provider's order, and if available utilize barcode medication administration as an additional check to ensure the correct dose is retrieved prior to administration. 2) Customers can inventory the CUBIE pocket via a cycle count or blind count to correct the inventory when discrepancies occur. Customers can refer to the BD Pyxis MedBank Cabinet User Guide - 10000414987 Ver. 01, Version v2.4 for instructions. Refer to chapter six (6), page 103. 3)Customers should ensure backup procedures (which may include but are not limited to keys, emergency medication supplies, alternate devices in proximity, etc) are in place and staff have all been properly educated on them. 4) Share this letter with any facilities within your network that you may have distributed affected devices to. 5) Post this Recall Letter on or near affected products as a reminder of the recall and the steps/instructions that are needed to be followed. 6) Complete and return the Response Form via email to [email protected] Firm is in the process of contacting all customers to schedule an appropriate time for the software upgrade to MedBank software version 1.7. If you require further assistance, contact the firm's technical support: 1-866-930-9251 Phone Hours: 24 hours, 7 days a week Website: https://bdx.my.site.com/CC360/s/impactedproducts?rn=MMS-24-5163/MMS-25-5183/5184
Quantity in commerce
7357
Distribution
US Nationwide distribution: CA, TX, WI, CO, AK, VA, TN, WA, NC, NY, MS, FL, KS, MI, IN, MN, MT, KY, MO, UT, NJ, OH, PA, ND, OR, NE, SD, NV, OK, GA, MA, IA, HI, ME, AZ, IL, SC, LA, WV, AL, ID, NM, DC, NH, VT, CT, MD, AR, WY.
Product codes and lots
REF/UDI-DI(GTIN)/Software Version: 169-143/10885403512568/2.4; 169-145/10885403512568/2.4; 169-144/10885403512568/2.4; 169-142/10885403512568/2.4; 139088-01/1.2.3.23; 155288-01/10885403508721/2.4; 169-137/10885403512544/2.4; 155311-01/10885403510847/2.4; 169-151/10885403512544/1.2.3.23; 138921-01/2.4; 155159-01/10885403504655/2.4; 169-101/10885403512544/2.4; 139122-01/2.4; 155160-01/10885403504662/2.4; 169-102/10885403512544/2.4; 138912-01/2.4; 155161-01/10885403504679/2.4; 169-103/10885403512544/2.4; 138902-01/2.4; 169-104/10885403512544/2.4; 138903-01/2.4; 155163-01/10885403504686/2.4; 169-105/10885403512544/2.4; 138908-01/2.4; 169-106/10885403512544/2.4; 138906-01/2.4; 155165-01/10885403504693/2.4; 169-107/10885403512544/2.4; 169-152/10885403512544/1.2.3.23; 138907-01/2.4; 155166-01/10885403504570/2.4; 169-108/10885403512544/2.4; 139131-01/2.4; 155167-01/10885403504587/2.4; 169-109/10885403512544/2.4; 139091-01/2.4; 155168-01/10885403504594/2.4; 169-110/10885403512544/2.4; 138909-01/2.4; 155169-01/10885403504600/2.4; 169-111/10885403512544/2.4; 138910-01/2.4; 155310-01/10885403510830/2.4; 169-112/10885403512544/2.4; 169-150/10885403512544/1.2.3.23; 138915-01/2.4; 155171-01/10885403504617/2.4; 169-113/10885403512544/2.4; 138943-01/2.4; 155081-01/10885403503870/2.4; 169-23/10885403512568/2.4; 138955-01/2.4; 155082-01/10885403503887/2.4; 169-24/10885403512568/2.4; 138924-01/2.4; 155083-01/10885403503894/2.4; 169-25/10885403512568/2.4; 138960-01/2.4; 155084-01/10885403503900/2.4; 169-26/10885403512568/2.4; 138961-01/2.4; 155085-01/10885403503917/2.4; 169-27/10885403512568/2.4; 138926-01/2.4; 155086-01/10885403503726/2.4; 169-28/10885403512568/2.4; 138927-01/2.4; 155087-01/10885403503733/2.4; 169-29/10885403512568/2.4; 155293-01/10885403509926/2.4; 169-138/10885403512568/2.4; 155303-01/10885403510779/2.4; 139082-01/2.4; 155088-01/10885403503740/2.4; 169-30/10885403512568/2.4; 155305-01/10885403510793/2.4; 139056-01/2.4; 155089-01/10885403503757/2.4; 169-31/10885403512568/2.4; 155304-01/10885403510786/2.4; 139045-01/2.4; 155090-01/10885403503764/2.4; 169-32/10885403512568/2.4; 138932-01/2.4; 155091-01/10885403503771/2.4; 169-33/10885403512568/2.4; 139084-01/2.4; 155092-01/10885403503788/2.4; 169-34/10885403512568/2.4; 138963-01/2.4; 155093-01/10885403503795/2.4; 169-35/10885403512568/2.4; 155294-01/10885403509933/2.4; 169-139/10885403512568/2.4; 138938-01/2.4; 155094-01/10885403503801/2.4; 169-36/10885403512568/2.4; 139125-01/2.4; 155095-01/10885403503818/2.4; 169-37/10885403512568/2.4; 138919-01/2.4; 155096-01/10885403503696/2.4; 169-38/10885403512568/2.4; 139126-01/2.4; 155097-01/10885403504198/2.4; 169-39/10885403512568/2.4; 139127-01/2.4; 155098-01/10885403503702/2.4; 169-40/10885403512568/2.4; 139043-01/2.4; 155099-01/10885403504204/2.4; 169-41/10885403512568/2.4; 138933-01/2.4; 155100-01/10885403504211/2.4; 169-42/10885403512568/2.4; 155299-01/10885403510526/2.4; 169-140/10885403512568/1.2.3.23; 139142-01/2.4; 155101-01/10885403504228/2.4; 169-43/10885403512568/2.4; 155302-01/10885403510762/2.4; 138959-01/2.4; 155102-01/10885403504235/2.4; 169-44/10885403512568/2.4; 138954-01/2.4; 155103-01/10885403504242/2.4; 169-45/10885403512568/2.4; 138934-01/2.4; 155104-01/10885403504259/2.4; 169-46/10885403512568/2.4; 138950-01/2.4; 155105-01/10885403503719/2.4; 169-47/10885403512568/2.4; 138947-01/2.4; 155106-01/10885403504266/2.4; 169-48/10885403512568/2.4; 138945-01/2.4; 155107-01/10885403504273/2.4; 169-49/10885403512568/2.4; 138931-01/2.4; 155108-01/10885403504280/2.4; 169-50/10885403512568/2.4; 138929-01/2.4; 155109-01/10885403504327/2.4; 169-51/10885403512568/2.4; 138946-01/2.4; 155110-01/10885403504297/2.4; 169-52/10885403512568/2.4; 138940-01/2.4; 155111-01/10885403504303/2.4; 169-53/10885403512568/2.4; 138952-01/2.4; 155112-01/10885403504099/2.4; 169-54/10885403512568/2.4; 138917-01/2.4; 155113-01/10885403504105/2.4; 169-55/10885403512568/2.4; 138918-01/2.4; 155114-01/10885403504112/2.4; 169-56/10885403512568/2.4; 138941-01/2.4; 155115-01/10885403504129/2.4; 169-57/10885403512568/2.4; 138905-01/2.4; 155116-01/10885403504136/2.4; 169-58/10885403512568/2.4; 138916-01/2.4; 155117-01/10885403504143/2.4; 169-59/10885403512568/2.4; 139141-01/2.4; 155118-01/10885403504150/2.4; 169-60/10885403512568/2.4; 138914-01/2.4; 155119-01/10885403504334/2.4; 169-61/10885403512568/2.4; 139063-01/2.4; 155120-01/10885403504167/2.4; 169-62/10885403512568/2.4; 138953-01/2.4; 155121-01/10885403504174/2.4; 169-63/10885403512568/2.4; 138936-01/2.4; 155122-01/10885403504181/2.4; 169-64/10885403512568/2.4; 138962-01/2.4; 155123-01/10885403504082/2.4; 169-65/10885403512568/2.4; 138944-01/2.4; 155124-01/10885403504075/2.4; 169-66/10885403512568/2.4; 138957-01/2.4; 155125-01/10885403503658/2.4; 169-67/10885403512568/2.4; 138922-01/2.4; 155126-01/10885403503665/2.4; 169-68/10885403512568/2.4; 138930-01/2.4; 155127-01/10885403503672/2.4; 169-69/10885403512568/2.4; 139128-01/2.4; 155128-01/10885403503689/2.4; 169-70/10885403512568/2.4; 138958-01/2.4; 155129-01/10885403504341/2.4; 169-71/10885403512568/2.4; 138920-01/2.4; 155130-01/10885403504358/2.4; 169-72/10885403512568/2.4; 138937-01/2.4; 155131-01/10885403504365/2.4; 169-73/10885403512568/2.4; 139086-01/2.4; 155132-01/10885403504372/2.4; 169-74/10885403512568/2.4; 138942-01/2.4; 155133-01/10885403503542/2.4; 169-75/10885403512568/2.4; 138949-01/2.4; 155134-01/10885403503559/2.4; 169-76/10885403512568/2.4; 139062-01/2.4; 155135-01/10885403503566/2.4; 169-77/10885403512568/2.4; 138939-01/2.4; 155136-01/10885403503573/2.4; 169-78/10885403512568/2.4; 138948-01/2.4; 155137-01/10885403503597/2.4; 169-79/10885403512568/2.4; 138956-01/2.4; 155138-01/10885403503603/2.4; 169-80/10885403512568/2.4; 138923-01/2.4; 155139-01/10885403503610/2.4; 169-81/10885403512568/2.4; 139163-01/2.4. 1119-00/10885403512704/2.4; 1139-00/10885403512704/2.4. 139054-01/2.4; 139055-01/2.4; 139053-01/2.4; 139041-01/2.4; 139046-01/2.4; 139038-01/2.4; 139040-01/2.4; 139049-01/2.4; 139044-01/2.4; 139039-01/2.4; 139051-01/2.4; 139164-01/2.4; 139052-01/2.4; 139070-01/2.4; 139071-01/2.4; 139061-01/2.4; 139066-01/2.4; 139065-01/2.4; 139058-01/2.4; 139060-01/2.4; 139067-01/2.4; 139064-01/2.4; 139059-01/2.4; 139068-01/2.4; 139165-01/2.4; 139069-01/2.4. 1144-00/2.4; 1137-00/2.4; 1147-00/10885403517181/2.4 1145-00/10885403517167/2.4

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1545-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1545-2025Class IIOngoingVoluntary: Firm initiated