Z-1583-2025Class II

Zenition 70. The device is to be used in health care facilities both inside and outside the operating room...

Philips Medical Systems Nederland B.V. / initiated 2023-07-20 / Open, Classified / root cause: software

Philips Medical Systems Nederland B.V. began a recall of Zenition 70. The device is to be used in health care facilities both inside and outside the operating room, sterile as well as non-sterile environment in a... on . FDA classified it as Class II and gives the reason as "Philips has become aware of the potential for unintended radiation exposure with Zenition 70 systems with Vascular Extension option and a Wireless Foot Switch". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-1583-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Philips Medical Systems Nederland B.V.
Product
Zenition 70. The device is to be used in health care facilities both inside and outside the operating room, sterile as well as non-sterile environment in a variety of procedures.
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
Philips has become aware of the potential for unintended radiation exposure with Zenition 70 systems with Vascular Extension option and a Wireless Foot Switch.
Root cause tag
software
FDA root cause
Software design
Action
Philips will send a customer notification to the existing users of the affected systems to alert them of the issue and provide the following actions that should be taken to prevent risks to patients: Keep the Electronic Product Radiation Correction letter with the documentation of the system. Where possible, place the Wireless Foot Switch within a 3 meter range from the C-arm stand to minimize as much as possible the potential delay of 72 milliseconds. Circulate this Electronic Product Radiation Correction letter to all users of the system so that they are aware of the issue. Philips will contact customers to schedule a visit to arrange for a Field Service Engineer visit to update the Software free of charge. (Philips reference 2024-IGT-PUN-007).
Quantity in commerce
12 systems
Distribution
US Nationwide distribution
Product codes and lots
Zenition 70

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1583-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1583-2025Class IIOngoingFDA Mandated