Zenition 70. The device is to be used in health care facilities both inside and outside the operating room...
Philips Medical Systems Nederland B.V. began a recall of Zenition 70. The device is to be used in health care facilities both inside and outside the operating room, sterile as well as non-sterile environment in a... on . FDA classified it as Class II and gives the reason as "Philips has become aware of the potential for unintended radiation exposure with Zenition 70 systems with Vascular Extension option and a Wireless Foot Switch". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1583-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Philips Medical Systems Nederland B.V.
- Product
- Zenition 70. The device is to be used in health care facilities both inside and outside the operating room, sterile as well as non-sterile environment in a variety of procedures.
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- Philips has become aware of the potential for unintended radiation exposure with Zenition 70 systems with Vascular Extension option and a Wireless Foot Switch.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Philips will send a customer notification to the existing users of the affected systems to alert them of the issue and provide the following actions that should be taken to prevent risks to patients: Keep the Electronic Product Radiation Correction letter with the documentation of the system. Where possible, place the Wireless Foot Switch within a 3 meter range from the C-arm stand to minimize as much as possible the potential delay of 72 milliseconds. Circulate this Electronic Product Radiation Correction letter to all users of the system so that they are aware of the issue. Philips will contact customers to schedule a visit to arrange for a Field Service Engineer visit to update the Software free of charge. (Philips reference 2024-IGT-PUN-007).
- Quantity in commerce
- 12 systems
- Distribution
- US Nationwide distribution
- Product codes and lots
- Zenition 70
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1583-2025 | Class II | Ongoing | FDA Mandated |