Z-1726-2025Class I

Welch Allyn Life2000 Compressor, Product Codes: 1) BT-80-0004; 2) BT-80-0004A; 3) BT-80-0008; 4) BT-80-0008A...

Baxter Healthcare Corporation / initiated 2025-04-07 / Open, Classified / root cause: software

Baxter Healthcare Corporation began a recall of Welch Allyn Life2000 Compressor, Product Codes: 1) BT-80-0004; 2) BT-80-0004A; 3) BT-80-0008; 4) BT-80-0008A; 5) RMS010093CP on . FDA classified it as Class I and gives the reason as "A cybersecurity vulnerability was discovered through internal testing". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-1726-2025
Classification
Class I
Status
Open, Classified
Initiated
Posted
Recalling firm
Baxter Healthcare Corporation
Product
Welch Allyn Life2000 Compressor, Product Codes: 1) BT-80-0004; 2) BT-80-0004A; 3) BT-80-0008; 4) BT-80-0008A; 5) RMS010093CP.
Product code
CBK Ventilator, Continuous, Facility Use
Reason for recall
A cybersecurity vulnerability was discovered through internal testing.
Root cause tag
software
FDA root cause
Software in the Use Environment
Action
Baxter issued an URGENT MEDICAL DEVICE RECALL notice to healthcare providers and patients beginning on 04/07/2025 via USPS first class mail. The notice explained that Baxter was voluntarily initiating a permanent removal of all Life 2000 ventilators currently in use due to a cybersecurity issue and requested the following: Actions to be Taken by DME Providers: 1. Notify your patients using Life2000 ventilators of this recall. 2. Work with patients and prescribers to identify an alternate therapy option. 3. Contact Baxter Acute Care Customer Service team at 800-426-4224, option 2, then option 1, to coordinate return of Life2000 ventilators and compressors. Actions to be Taken by Healthcare Providers: 1. Locate and discontinue use of all Life2000 ventilators and compressors within your facility. 2. Contact Baxter Acute Care Customer Service team at 800-426-4224, option 2, then option 1, to coordinate return of Life2000 ventilators and compressors at your facility. Actions to be Taken by Distributors: 1. Please disseminate this information and conduct a user-level recall of the affected product that you distributed to customers. 2. Contact Baxter Acute Care Customer Service team at 800-426-4224, option 2, then option 1, to coordinate return of Life2000 ventilators and compressors. If you suspect that a device has been exposed to unauthorized personnel, contact Baxter Acute Care Customer Service team. The team can help verify whether the device settings are correct and if the device is performing as expected. The team can be reached by email at [email protected], or by phone at 800-426-4224, option 2, then option 1. For questions regarding this communication, contact Baxter Acute Care Customer Service team at 800-426-4224, option 2, then option 1, between the hours of 8:00 am and 6:00 pm Central Time, Monday through Friday.
Quantity in commerce
4174 units
Distribution
Worldwide distribution.
Product codes and lots
ALL SERIAL NUMBERS: 1) BT-80-0004, UDI/DI 00887761978041 or 00815410020292 or 00887761978072; 2) BT-80-0004A, UDI/DI 00887761978041 or 00815410020292 or 00887761978072; 3) BT-80-0008, UDI/DI 00887761976283; 4) BT-80-0008A, UDI/DI 00887761976283; 5) RMS010093CP, UDI/DI 00887761978041 or 00815410020537 or 00815410020292.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1726-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1725-2025Class IOngoingVoluntary: Firm initiated
Z-1726-2025Class IOngoingVoluntary: Firm initiated