HistoCore PELORIS 3, Model/Catalog Number: 45.0005/45.7512.501 A11, Software Version: 3.4.0. The HistoCore...

Leica Biosystems Melbourne Pty Ltd / initiated 2025-04-02 / Open, Classified

Leica Biosystems Melbourne Pty Ltd began a recall of HistoCore PELORIS 3, Model/Catalog Number: 45.0005/45.7512.501 A11, Software Version: 3.4.0. The HistoCore PELORIS 3 Rapid Tissue Processor is a dual retort... on . FDA classified it as Class II and gives the reason as "There is a leakage issue associated with the tubing in the manifold of the instrument". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-1683-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Leica Biosystems Melbourne Pty Ltd
Product
HistoCore PELORIS 3, Model/Catalog Number: 45.0005/45.7512.501 A11, Software Version: 3.4.0. The HistoCore PELORIS 3 Rapid Tissue Processor is a dual retort rapid tissue processor used to prepare tissue samples.
Product code
IEO Processor, Tissue, Automated
Reason for recall
There is a leakage issue associated with the tubing in the manifold of the instrument.
FDA root cause
Release of Material/Component prior to receiving test results
Action
An URGENT MEDICAL DEVICE RECALL notification letter dated 3/31/25 was sent to customers. ACTIONS REQUIRED 1. If leakage from the instrument is observed in the drip tray or surrounding the instrument, please follow your laboratory's leakage containment/safety protocol. If a protocol is running, wait for the run to finish, switch off the instrument and contact your local Leica Biosystems representative immediately. 2. If no leakage has been observed from the instrument, continue using it as normal. Conduct routine monitoring for leakage until the on-site inspection is completed. 3. Your local Leica Biosystems representative will contact you to schedule an onsite inspection of your instruments tubing and reseat the tubes if needed. 4. Confirm Receipt of Recall Notice Please confirm receipt of this notice by signing and dating the attached Recall Acknowledgement Form by email to the following email address: lbsme/[email protected] Please contact your local Leica Biosystems representative immediately if you have any questions or concerns. An URGENT MEDICAL DEVICE RECALL notification letter dated 8/13/25 was sent to customers. ACTIONS REQUIRED 1. If leakage from the instrument is observed in the drip tray or surrounding the instrument, please follow your laboratory's leakage containment/safety protocol. If a protocol is running, wait for the run to finish, switch off the instrument and contact your local Leica Biosystems representative immediately. 2. If no leakage has been observed from the instrument, continue using it as normal. Conduct routine monitoring for leakage until the on-site inspection is completed. 3. Your local Leica Biosystems representative will contact you to schedule an onsite inspection of your instrument(s) for replacing the tubing and/or density meter manifold if needed. 4. Confirm Receipt of Recall Notice Please confirm receipt of this notice by signing and dating the attached Recall Acknowledgement Form by email to the following email ad
Quantity in commerce
77 units
Distribution
US States: AZ, CA, FL, MN, NJ, NY, NC, OH, PA, TN, TX, UT, VA, and WI.
Product codes and lots
UDI-DI: 09349458004811; Serial Numbers: 45111117, 45111154, 45111155, 45111156, 45111157, 45111158, 45111159, 45111160, 45111161, 45111162, 45111164, 45111166, 45111167, 45111168, 45111169, 45111170, 45111171, 45111174, 45111175, 45111176, 45111177, 45111178, 45111179, 45111180, 45111181, 45111182, 45111183. 08/12/2025 UPDATE: Additional Serial Numbers: 45111064, 45111065, 45111067, 45111068, 45111069, 45111070, 45111071, 45111073, 45111074, 45111075, 45111076, 45111077, 45111078, 45111079, 45111080, 45111081, 45111082, 45111084, 45111086, 45111087, 45111088, 45111089, 45111090, 45111091, 45111092, 45111093, 45111094, 45111095, 45111097, 45111098, 45111099, 45111100, 45111101, 45111102, 45111103, 45111104, 45111108, 45111109, 45111110, 45111111, 45111112, 45111113, 45111114, 45111115, 45111116, 45111120, 45111122, 45111123, 45111124, 45111125, 45111126, 45111127, 45111128, 45111129, 45111130, 45111131, 45111132, 45111133, 45111134, 45111135, 45111136, 45111137, 45111139, 45111140, 45111141, 45111142, 45111143, 45111144, 45111145, 45111146, 45111147, 45111148, 45111149, 45111150, 45111151, 45111152, 45111163.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1683-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1683-2025Class IIOngoingVoluntary: Firm initiated