Z-1681-2025Class I

CADD-Solis Infusion Pump Models: 1) PUMP KIT, CADD-SOLIS, MDL 2110, V4.2, CE ENGLISH 1/EA, Model Number...

Abbott Medical / initiated 2025-04-10 / Open, Classified

Smiths Medical ASD, Inc. began a recall of CADD-Solis Infusion Pump Models: 1) PUMP KIT, CADD-SOLIS, MDL 2110, V4.2, CE ENGLISH 1/EA, Model Number: 21-2111-0402-51; 2) PUMP KIT, CADD-SOLIS, MDL 2110... on . FDA classified it as Class I and gives the reason as "There is a potential for thermal damage in CADD-Solis and CADD- Solis VIP infusion pumps". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-1681-2025
Classification
Class I
Status
Open, Classified
Initiated
Posted
Recalling firm
Abbott Medical
Product
CADD-Solis Infusion Pump Models: 1) PUMP KIT, CADD-SOLIS, MDL 2110, V4.2, CE ENGLISH 1/EA, Model Number: 21-2111-0402-51; 2) PUMP KIT, CADD-SOLIS, MDL 2110, V4.2, YELLOW, CE ENGLISH 1/EA, Model Number: 21-2112-0402-51; 3) PUMP KIT, CADD-SOLIS, YELLOW, MODEL 2110 1/EA, Model Number: 21-2112-0100-51; 4) PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW KEYPAD, MODEL 2110, NA ENGLISH 1/EA, Model Number: 21-2112-0300-01; 5) PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, NA ENGLISH 1/EA, Model Number: 21-2111-0300-01; 6) PUMP KIT, CADD-SOLIS, MDL 2110, V4.1, CE ENGLISH 1/EA, Model Number: 21-2111-0401-51; 7) PUMP KIT, CADD-SOLIS, MODEL 2110, CE ENGLISH 1/EA, Model Number: 21-2111-0100-51; 8) PUMP KIT, CADD-SOLIS, MDL 2110, V4.1, YELLOW, CE ENGLISH 1/EA, Model Number: 21-2112-0401-51; 9) PUMP KIT, CADD-SOLIS, MDL 2100, CE ENGLISH 1/EA, Model Number: 21-2101-51; 10) PUMP KIT, CADD-SOLIS, MDL 2110, V4.0, CE ENGLISH 1/EA, Model Number: 21-2111-0400-51; 11) PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, OUS ENGLISH 1/EA, Model Number: 21-2111-0300-50; 12) PUMP, TRIAL/LOANER, CADD- SOLIS, MODEL 2100 1/EA, Model Number: 21-2101-00; 13) PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110 TRIAL/LOANER, NA ENGLISH 1/EA, Model Number: 21-2111-0300-00; 14) PUMP KIT, CADD-SOLIS, MDL 2110, TRIAL/LOANER, V4.2, CE ENGLISH 1/EA, Model Number: 21-2111-0402-78; 15) PUMP, TRIAL/LOANER, CADD- SOLIS, MODEL 2110 1/EA, Model Number: 21-2111-0100-00; 16) PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110,TRIAL/LOANER, NA ENGLISH 1/EA, Model Number: 21-2112-0300-00; 17) PUMP KIT, CADD-SOLIS, MDL 2110, V4.2, TRIAL/LOANER YELLOW, CE ENGLISH 1/EA, Model Number: 21-2112-0402-78; 18) PUMP, DEMO, CADD-SOLIS, MODEL 2100, NOT FOR HUMAN USE 1/EA, Model Number: 21-2101-249; 19) PUMP KIT, CADD-SOLIS, MDL 2110, TRIAL/LOANER, V4.1, CE ENGLISH 1/EA, Model Number: 21-2111-0401-78; 20) PUMP KIT, CADD-SOLIS, YELLOW, MDL 2100 1/EA, Model Number: 21-2102-51; 21) PUMP KIT, CADD-SOLIS, MDL 2110, V4.1, NFHU, YELLOW, CE ENGLISH 1/EA, Model Number: 21-2112-0401-249; 22) PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110, NFHU, NA ENGLISH 1/EA, Model Number: 21-2111-0300-231; 23) PUMP KIT, CADD-SOLIS, MDL 2110, V4.1, TRIAL/LOANER YELLOW, CE ENGLISH 1/EA, Model Number: 21-2112-0401-78; 24) PUMP KIT, CADD-SOLIS, MDL 2110, V4.0, YELLOW, CE ENGLISH 1/EA, Model Number: 21-2112-0400-51; 25) PUMP KIT, CADD-SOLIS HPCA PIB, YELLOW, MODEL 2110, OUS ENGLISH 1/EA, Model Number: 21-2112-0300-50; 26) PUMP KIT, CADD-SOLIS HPCA PIB, MODEL 2110 TRIAL/LOANER, NA ENGLISH 1/EA, Model Number: 21-2111-0300-00; 27) PUMP, TRIAL/LOANER, CADD- SOLIS, MODEL 2110 1/EA, Model Number: 21-2111-0100-00; 28) PUMP KIT, CADD-SOLIS, MODEL 2110, CE ENGLISH 1/EA, Model Number: 21-2111-0100-51; 29) PUMP, CADD-SOLIS, MODEL 2110, CE ENGLISH OUS 1/EA, Model Number: 21-2111-0100-50; 30) PUMP KIT, CADD-SOLIS, MDL 2100, CE ENGLISH 1/EA, Model Number: 21-2101-51; 31) PUMP KIT, CADD-SOLIS HPCA, YELLOW KEYPAD, MODEL 2110, CE ENGLISH, OUS 1/EA, Model Number: 21-2112-0100-50; 32) PUMP, TRIAL/LOANER, CADD- SOLIS, MODEL 2100 1/EA, Model Number: 21-2101-00; 33) PUMP KIT, CADD-SOLIS, MDL 2110, v4.0, YELLOW, ENGLISH 1/EA, Model Number: 21-2112-0400-01.
Product code
MEA Pump, Infusion, Pca
Reason for recall
There is a potential for thermal damage in CADD-Solis and CADD- Solis VIP infusion pumps.
FDA root cause
Under Investigation by firm
Action
An URGENT MEDICAL DEVICE CORRECTION dated 4/10/25 was sent to customers. Actions for Users: Inform all affected CADD-Solis and CADD-Solis VIP users of this notice. Provide the instructions below: 1. Ensure all users or potential users of these products are immediately made aware of this notification. 2. Upon receipt of this notification and ongoing during regular maintenance, visually inspect the external condition of the battery pack or Wireless Communication Module and look for evidence of any physical damage. Handle a damaged device with caution. As stated in the battery pack Instructions for Use, if the battery pack housing is cracked or otherwise damaged, replace the battery pack. NEVER use a battery pack that appears damaged. A rechargeable battery pack must be replaced with either another CADD-Solis rechargeable battery pack or with 4 AA batteries. 3. Visually inspect the condition of the battery compartment for evidence of any physical damage. Contact Technical Support for repairs. 4. Always check the battery compartment for fluid or debris before inserting the batteries, and do not allow any fluid or debris to fall into the battery compartment. 5. Complete the electronic Response Form via the link provided within this email notification within ten days of receipt to acknowledge your understanding of this notification, even if you do not have the affected product. 6. DISTRIBUTORS: If you have distributed potentially affected products to your customers, please immediately forward your customer list to Sedgwick at [email protected]. Sedgwick will notify customers with email addresses via email and will send physical notifications to customers without email addresses in order for them to complete the Response Form Follow up Actions by Smiths Medical: Smiths Medical is continuing to investigate this matter to determine if additional actions may be warranted. For further inquiries, please contact Smiths Medical using the following information: Smit
Quantity in commerce
313,676 total units
Distribution
Worldwide - US Nationwide distribution.
Product codes and lots
All serial numbers. 1 ) 21-2111-0402-51 , UDI-DI: 15019517154290 ; 2 ) 21-2112-0402-51 , UDI-DI: 15019517154993 ; 3 ) 21-2112-0100-51 , UDI-DI: 10610586039232 ; 7 ) 21-2111-0100-51 , UDI-DI: 30610586039205 ; 9 ) 21-2101-51 , UDI-DI: 10610586031854 ; 13 ) 21-2111-0300-00 , UDI-DI: 10610586038792 ; 14 ) 21-2111-0402-78 , UDI-DI: 15019517154887 ; 15 ) 21-2111-0100-00 , UDI-DI: 10610586039195 ; 16 ) 21-2112-0300-00 , UDI-DI: 10610586038822 ; 17 ) 21-2112-0402-78 , UDI-DI: 15019517155013 ; 18 ) 21-2101-249 , UDI-DI: 10610586034787 ; 20 ) 21-2102-51 , UDI-DI: 10610586033933 ; 22 ) 21-2111-0300-231 , UDI-DI: 10610586038785 ; 23 ) 21-2112-0401-78 , UDI-DI: 15019517101034 ; 24 ) 21-2112-0400-51 , UDI-DI: 10610586041051 ; 25 ) 21-2112-0300-50 , UDI-DI: 10610586038860 ; 26 ) 21-2111-0300-00 , UDI-DI: 10610586038792 ; 27 ) 21-2111-0100-00 , UDI-DI: 10610586039195 ; 28 ) 21-2111-0100-51 , UDI-DI: 30610586039205 ; 29 ) 21-2111-0100-50 , UDI-DI: 10610586039256 ; 30 ) 21-2101-51 , UDI-DI: 10610586031854 ; 31 ) 21-2112-0100-50 , UDI-DI: 10610586039270 ; 32 ) 21-2101-00 , UDI-DI: 35019517154904; 33 ) 21-2112-0400-01 , UDI-DI: 10610586041822 ;

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1681-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1681-2025Class IOngoingVoluntary: Firm initiated
Z-1682-2025Class IOngoingVoluntary: Firm initiated