Z-1739-2025Class II
Atalante X, Powered Exoskeleton. Atalante X is intended to perform ambulatory functions and mobility...
Wandercraft / initiated 2025-04-03 / Open, Classified
WANDERCRAFT SAS began a recall of Atalante X, Powered Exoskeleton. Atalante X is intended to perform ambulatory functions and mobility exercises, hands-free, in rehabilitation institutions... on . FDA classified it as Class II and gives the reason as "Potential for exoskeleton to lose lateral balance and cause patient to fall". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-1739-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Wandercraft
- Product
- Atalante X, Powered Exoskeleton. Atalante X is intended to perform ambulatory functions and mobility exercises, hands-free, in rehabilitation institutions under the supervision of a trained operator.
- Product code
- PHL Powered Exoskeleton
- Reason for recall
- Potential for exoskeleton to lose lateral balance and cause patient to fall.
- FDA root cause
- Use error
- Action
- An URGENT SAFETY NOTICE dated 4/3/25 was mailed to consignees. The letter notifies consignees that WANDERCRAFT will be reinforcing their training program for the Atalante X and providing additional instructions by 5/1/25. Until the additional instructions can be provided, users are being instructed to only have certified Atalante X operators maintain the stability of the exoskeleton at all times by keeping at least one hand on the handles throughout the session. Consignees are to share the recall notification with users at their facility and retained for awareness.
- Quantity in commerce
- 3
- Distribution
- Domestic: PA & NY.
- Product codes and lots
- UDI: (01)3665965000129(21)V5_ABP(11)AA06JJ, (01)3665965000129(21)V5_ACH(11)AA09JJ, (01)3665965000129(21)V5_ADO(11)AA02JJ. No model or serial number.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1739-2025 | Class II | Ongoing | Voluntary: Firm initiated |