Z-1702-2025Class II

Tibial Augment; Item Numbers: (1) 880-331/11 (Sz. 1-2 - cemented x-sm), (2) 880-331/12 (Sz. 1-2 - cemented...

Waldemar Link GmbH & Co. KG / initiated 2025-04-03 / Open, Classified / root cause: manufacturing process

Waldemar Link GmbH & Co. KG (Mfg Site) began a recall of Tibial Augment; Item Numbers: (1) 880-331/11 (Sz. 1-2 - cemented x-sm), (2) 880-331/12 (Sz. 1-2 - cemented sm), (3) 880-331/13 (Sz. 1-2 - cemented md), (4)... on . FDA classified it as Class II and gives the reason as "The device was delivered with a preassembled Femoral Augment screw that was missing its thread". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1702-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Waldemar Link GmbH & Co. KG
Product
Tibial Augment; Item Numbers: (1) 880-331/11 (Sz. 1-2 - cemented x-sm), (2) 880-331/12 (Sz. 1-2 - cemented sm), (3) 880-331/13 (Sz. 1-2 - cemented md), (4) 880-331/21 (Sz. 1-2 - cemented lg), (5) 880-331/22 (Sz. 1-2 - cemented x-lg), (6) 880-331/23 (Sz. 1-2 - cemented xx-lg), (7) 880-333/11 (Sz. 3-4 - cemented x-sm), (8) 880-333/12 (Sz. 3-4 - cemented sm), (9) 880-333/13 (Sz. 3-4 - cemented md), (10) 880-333/21 (Sz. 3-4 - cemented lg), (11) 880-333/22 (Sz. 3-4 - cemented x-lg), (12) 880-333/23 (Sz. 3-4 - cemented xx-lg), (13) 880-335/11 (Sz. 5-6 - cemented x-sm), (14) 880-335/12 (Sz. 5-6 - cemented sm), (15) 880-335/13 (Sz. 5-6 - cemented md), (16) 880-335/21 (Sz. 5-6 - cemented lg), (17) 880-335/22 (Sz. 5-6 - cemented lg), (18) 880-335/23 (Sz. 5-6 - cemented xx-lg), (19) 880-337/11 (Sz. 7-8 - cemented x-sm), (20) 880-337/12 (Sz. 5-6 - cemented sm), (21) 880-337/13 (Sz. 7-8 - cemented md), (22) 880-337/21 (Sz. 7-8 - cemented lg), (23) 880-337/22 (Sz. 7-8 - cemented x-lg), (24) 880-337/23 (Sz. 7-8 - cemented xx-lg), (25) 880-339/11 (Sz. 9-10 - cemented x-sm), (26) 880-339/12 (Sz. 9-10 - cemented sm), (27) 880-339/13 (Sz. 9-10 - cemented md), (28) 880-339/21 (Sz. 9-10 - cemented lg), (29) 880-339/22 (Sz. 9-10 - cemented x-lg), (30) 880-339/23 (Sz. 9-10 - cemented xx-lg);
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
The device was delivered with a preassembled Femoral Augment screw that was missing its thread.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
On April 3, 2025 URGENT: MEDICAL DEVICE CORRECTION letters were emailed to customers. Actions to be taken by the Customer/User: Please take note of the following instructions for replacing any screws found missing the thread. No product return is required. Instructions for disassembling and replacing a defective screw: Disassembly 1. Use the 2.5 mm Torque Wrench article REF (15-2545) to unscrew the screw from the Augment while pushing the tip of the screw towards the Augment. 2. Exchange the screw. Open an alternative Augment and use its screw. Alternative sizes of Augments are always available in the operating room during the surgery. For all Femoral Augment sizes and types the same long screw type is preassembled and for all Tibial Augments the same short screw type is preassembled, see table 1. Assembly/Replacement 3. Use the 2.5 mm Torque Wrench article REF (15-2545) to screw in the replacement screw into the Augment. 4. The screw head must be countersunk in the Augment and the screw tip must protrude. Follow the LinkSymphoKnee surgical technique for all other steps. Please respond using the Distributor Reply form to indicate your understanding of this Device Correction and of the need to apply the instructions provided above to replace any affected screws. Type of Action by the Company: LINK will provide sterilized replacement screws (for both Femoral Augments and Tibial Augments) for each affected distributor as a back-up for upcoming surgeries. Other Information: Opening an additional Augment to exchange the screw will not incur any costs to you. Should you have any questions about acquiring backup/replacement screws for forthcoming surgeries (once available), please contact Customer Service at LinkBio ([email protected]). Please contact LinkBio at 973-625-1333 or [email protected] for any additional questions.
Quantity in commerce
1771 units
Distribution
US Nationwide distribution in the states of AL, AZ, CA, CO, FL, GA, IN, IL, KS, LA, MD, MO, NV, OH, TX, VA, WI.
Product codes and lots
Item Number/UDI-DI: (1) 880-331/11 (04026575258192), (2) 880-331/12 (04026575258208), (3) 880-331/13 (04026575258215), (4) 880-331/21 (04026575258222), (5) 880-331/22 (04026575258246), (6) 880-331/23 (04026575258253), (7) 880-333/11 (04026575258260), (8) 880-333/12 (04026575258277), (9) 880-333/13 (04026575258284), (10) 880-333/21 (04026575258291), (11) 880-333/22 (04026575258307), (12) 880-333/23 (04026575258314), (13) 880-335/11 (04026575258321), (14) 880-335/12 (04026575258338), (15) 880-335/13 (04026575258345), (16) 880-335/21 (04026575258352), (17) 880-335/22 (04026575258369), (18) 880-335/23 (04026575258376), (19) 880-337/11 (04026575258383), (20) 880-337/12 (04026575258390), (21) 880-337/13 (04026575258406), (22) 880-337/21 (04026575258413), (23) 880-337/22 (04026575258420), (24) 880-337/23 (04026575258437), (25) 880-339/11 (04026575258475), (26) 880-339/12 (04026575258505), (27) 880-339/13 (04026575258529), (28) 880-339/21 (04026575258536), (29) 880-339/22 (04026575258543), (30) 880-339/23 (04026575258550); Applies to all lots of the LINK SymphoKnee Femoral Augments listed, manufactured up until 03/01/2025.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1702-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1699-2025Class IIOngoingVoluntary: Firm initiated
Z-1700-2025Class IIOngoingVoluntary: Firm initiated
Z-1701-2025Class IIOngoingVoluntary: Firm initiated
Z-1702-2025Class IIOngoingVoluntary: Firm initiated