Bard Minnesota Four Lumen Esophagogastric Tamponade Tube, REF:0092220 18 Fr. (6mm, 7/32"), 41" Length (104cm)...
C.R. Bard Inc began a recall of Bard Minnesota Four Lumen Esophagogastric Tamponade Tube, REF:0092220 18 Fr. (6mm, 7/32"), 41" Length (104cm) with 8" (20.3cm) Balloon X-Ray Opaque Rubber on . FDA classified it as Class I and gives the reason as "Due to customer complaints, there is a potential for users to be unable to or find it difficult to remove the plastic plugs in order to inflate the gastric and/or esophageal balloons". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1755-2025
- Classification
- Class I
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- C.R. Bard Inc
- Product
- Bard Minnesota Four Lumen Esophagogastric Tamponade Tube, REF:0092220 18 Fr. (6mm, 7/32"), 41" Length (104cm) with 8" (20.3cm) Balloon X-Ray Opaque Rubber
- Product code
- KNT Tubes, Gastrointestinal (And Accessories)
- Reason for recall
- Due to customer complaints, there is a potential for users to be unable to or find it difficult to remove the plastic plugs in order to inflate the gastric and/or esophageal balloons.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On April 17, 2025, BD (C.R. Bard) Urology and Critical Care issued a "Urgent Medical Device Correction" Notification to affected consignees. BD asked consignees to take the following actions: 1. Follow the above plastic plug removal instructions. 2. Post this notice where the devices are stored. 3. Please check all inventory locations within your institution for affected products 4. Share this notice with any users of the product within your facilities or with any interfacility users where product was transferred 5.Complete and return the attached Customer Response Form to the BD contact noted on the form confirming acknowledgement of this notification 6.Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program via: Web: MedWatch website at www.fda.gov/medwatch Phone: 1-800-FDA-1088 (1-800-332-1088)
- Distribution
- US: AK AL AR AZ CA CO CT DC DE FL GA HI IA ID IL IN KS KY LA MA MD ME MI MN MO MS MT NC ND NE NH NJ NM NV NY OH OK OR PA RI SC SD TN TX UT VA VT WA WI WV WY, Puerto Rico OUS: Singapore, Canada, Belgium, Ireland, Netherlands, South Africa, Sweden, United Kingdom
- Product codes and lots
- All Lots within expiration/UDI: (01)00801741076824
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1755-2025 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-1756-2025 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-1757-2025 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-1758-2025 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-1759-2025 | Class I | Ongoing | Voluntary: Firm initiated |