Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1785-2025Class II

True 3D Viewer, Model: EP-00003

Echopixel, Inc. / initiated 2018-02-21 / Open, Classified / root cause: design

Echopixel, Inc. began a recall of True 3D Viewer, Model: EP-00003 on . FDA classified it as Class II and gives the reason as "With certain software versions, import of enhanced ultrasound formatted data/ultrasound images into 3D viewer, (for digital image processing/review/analysis/communication/media interchange acquired from...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-1785-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Echopixel, Inc.
Product
True 3D Viewer, Model: EP-00003
Product code
LLZ System, Image Processing, Radiological
Reason for recall
With certain software versions, import of enhanced ultrasound formatted data/ultrasound images into 3D viewer, (for digital image processing/review/analysis/communication/media interchange acquired from CT/MRI/ultrasound, and pre-operative analysis of surgical options), can result in incorrect display of image orientation, which could result in misdiagnosis or incorrect surgical treatment plan.
Root cause tag
design
FDA root cause
Device Design
Action
On 2/21/2018, correction notices were emailed to customers who were asked to do the following: a) Affected software versions should not be used with Enhanced 3D Ultrasound DICOM compliant volumes or Philips 3DDCM. b) If clinicians are utilizing Enhanced 3D Ultrasound DICOM compliant volumes or Philips 3DDCM volume data sets, verify, retrospectively, any potential risk to health to patients and further determine if patients need to be reevaluated and report this information to the firm. c) Pass a copy of this notice to those staff members who need to be aware of the issue and maintain awareness for a three-month period. d)If devices were transferred to other departments or organizations, provide a copy of this notice to these consignees. Firm will contact customers to provide information about the timing and delivery mechanisms of the software upgrade. If you have questions, contact the firm at: 1-844-273-7766 x2
Quantity in commerce
18
Distribution
Worldwide - US Nationwide distribution in the states of CA, OH, NJ, PA, MA, FL, CT, TX, LA, MI, MN and the country of CA.
Product codes and lots
UDI-DI: 00851325007003. Full release software versions: 1.6.1.3080.428, released 13-Apr-2017 1.6.2.3670.435, released 07-Dec-2017 Develop Versions, for research use: 0.00.2695.422, released 15-Feb-2017 0.00.2927.423, released 15-Feb-2017 0.00.2964.423, released 24-Feb-2017 0.00.3102.428, released 04-Apr-2017 0.00.3245.428, released 05-May-2017 0.00.3302.431, released 02-Jun-2017 0.00.3344.432, released 18-Jul-2017 0.00.3412.433, released 14-Aug-2017 0.00.3656.442, released 09-Nov-2017 0.00.3669.442, released 05-Dec-2017 0.00.3683.445, released 02-Jan-2018 0.00.3702.447, released 11-Jan-2018 0.00.3730.447, released 08-Feb-2018

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1785-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1785-2025Class IIOngoingVoluntary: Firm initiated