Z-1666-2025Class II

Tempus Pro Patient Monitor, REFs: 00-1004-R (Tempus Pro non-printer, IBP, without headset dock), 00-1007-R...

Remote Diagnostic Technologies Ltd. / initiated 2025-04-16 / Open, Classified / root cause: design

Remote Diagnostic Technologies Ltd. began a recall of Tempus Pro Patient Monitor, REFs: 00-1004-R (Tempus Pro non-printer, IBP, without headset dock), 00-1007-R and 00-1007 (Tempus Pro non-printer, IBP, with... on . FDA classified it as Class II and gives the reason as "software issue that causes an error screen on a portable vital signs monitor when attaching video Laryngoscope". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-1666-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Remote Diagnostic Technologies Ltd.
Product
Tempus Pro Patient Monitor, REFs: 00-1004-R (Tempus Pro non-printer, IBP, without headset dock), 00-1007-R and 00-1007 (Tempus Pro non-printer, IBP, with headset dock), 00-1024-R (Tempus Pro printer), 00-1026-R (Tempus Pro printer, IBP), 00-1007 [(Tempus Pro, BT Dock (IB Only)], 00-1004 00-1004 [Tempus Pro (IB Only)]
Product code
MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Reason for recall
software issue that causes an error screen on a portable vital signs monitor when attaching video Laryngoscope.
Root cause tag
design
FDA root cause
Device Design
Action
On April 16, 2025, Remote Diagnostic Technologies Ltd (subsidiary of Philips) issued a Urgent Medical Device Correction to affected consignees via mail. Philips ask consignees to take the following: 1. DO NOT connect the Video Laryngoscope USB probe to the Tempus Pro during device boot-up. Once Patient Details dialog appears, enter patient data, then plug in the Video Laryngoscope USB probe. 2. If the Tempus Pro Monitor displays the error dialog screen (where the user is asked to restart the device), press the Shutdown Button and restart the Tempus Pro Monitor. 3. Keep a copy of this letter with or near your Tempus Pro Monitor. 4. Complete and return the Urgent Field Safety Notification Response Form included with this letter within 30 days of receipt of this notice. 5. This notice must be shared with all relevant personnel within your organization and with any organization where the potentially affected devices have been transferred.
Quantity in commerce
6,961 units
Distribution
Worldwide - US Nationwide distribution including in the states of AL, AZ, CA, CO, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, WI, WV, WY and the countries of AE, AF, AT, AU, BE, BO, CA, CH, CZ, DE, DK, ES, FI, FR, GB, HK, IE, IL, IS, IT, KY, MY, NL, NO, OM, PL, PT, RO, SA, SE, SG, SI, SL, TH, ZA.
Product codes and lots
All Lots/ UDI-DI(s): 00-1004-R - 05060472440020 & 05060472442901; 00-1007-R - 05060472440013 & 05060472442918; 00-1024-R - 05060472441027 & 05060472442925; 00-1026-R - 05060472441058 & 05060472442932; 00-1004 - 05060472440624; 00-1007 - 05060472440655

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1666-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1666-2025Class IIOngoingVoluntary: Firm initiated