Z-1821-2025Class II

Atellica IM Homocysteine Calibrator (2 Pack). Siemens Material Number: 10995498

Siemens Healthcare Diagnostics Inc. / initiated 2025-04-15 / Open, Classified / root cause: labeling and UDI

Siemens Healthcare Diagnostics, Inc. began a recall of Atellica IM Homocysteine Calibrator (2 Pack). Siemens Material Number: 10995498 on . FDA classified it as Class II and gives the reason as "Certain lots of Atellica IM Homocysteine Calibrator and ADVIA Centaur HCY Calibrator do not meet the assay Instructions for Use (IFU) reference interval due to calibrator instability resulting in increased positive bias". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1821-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Atellica IM Homocysteine Calibrator (2 Pack). Siemens Material Number: 10995498
Product code
JIS Calibrator, Primary
Reason for recall
Certain lots of Atellica IM Homocysteine Calibrator and ADVIA Centaur HCY Calibrator do not meet the assay Instructions for Use (IFU) reference interval due to calibrator instability resulting in increased positive bias.
Root cause tag
labeling and UDI
FDA root cause
Under Investigation by firm
Action
An Urgent Medical Device Correction (UMDC) was mailed to all US affected customers by FedEx, and an Urgent Field Safety Notice (UFSN) was issued to all OUS Siemens reporting sites with affected customers on about 04/15/2025. Customers were instructed to review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, discontinue use of and discard the affected devices, and complete and return the Effectiveness Check questionnaire. Customers were also instructed to forward the letter to those who may have received the affected devices.
Quantity in commerce
2,212 packs (4,424 units) - 448 packs US, 1,764 packs OUS
Distribution
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KY, MA, MD, ME, MI, MO, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, TX, UT, VA, VT, WI, WY and the countries of Austria, Belgium, Croatia, Curacao,¿St¿Eus, Czech¿Republic, Denmark, Egypt, Finland, France, Georgia, Germany, Greece, Hungary, Iran, Ireland, Israel, Italy, Jordan, Kuwait, Latvia, Lithuania, Nepal, Netherlands, Norway, Pakistan, Poland, Portugal, Romania, Serbia, Slovakia, Spain, Sweden, Switzerland, Turkey, Turkmenistan, U.A.E., United¿Kingdom, Vatikan City, Zaire.
Product codes and lots
Siemens Material Number: 10995498. UDI Number: (01)00630414598253(10)86236A74(17)20250514, (01)00630414598253(10)44206A75(17)20250925. Kit lot numbers: 86236A74, 44206A75.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1821-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1821-2025Class IIOngoingVoluntary: Firm initiated
Z-1822-2025Class IIOngoingVoluntary: Firm initiated