Z-0033-2026Class II
Medtronic, Simplera Sensor, REF: MMT-5100JD1
Medtronic, Inc. / initiated 2025-05-07 / Open, Classified / root cause: manufacturing process
Medtronic MiniMed, Inc. began a recall of Medtronic, Simplera Sensor, REF: MMT-5100JD1 on . FDA classified it as Class II and gives the reason as "The device cap label may not adhere to the sensor, compromising the sterility of the sensor, and increasing the risk of infection". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0033-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Medtronic, Inc.
- Product
- Medtronic, Simplera Sensor, REF: MMT-5100JD1
- Product code
- MDS Sensor, Glucose, Invasive
- Reason for recall
- The device cap label may not adhere to the sensor, compromising the sterility of the sensor, and increasing the risk of infection.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- On May 7, 2025, Medtronic issued a "Urgent Field Safety Notice" to affected consignees. Medtronic asked consignees to take the following actions: 1. Verify if the sensors you currently have are from the impacted lot. 2. If you have a Simplera" sensor from the impacted lot, do not use this Simplera" sensor and dispose of it according to local regulations. 3. Please contact our helpline/your Medtronic contact at 01923 205167 to request a replacement. 4. Please acknowledge that you have read and understood this notification and have followed the actions listed in this letter, even if your sensor is unaffected, or if you do not have any sensors from the affected lot.
- Quantity in commerce
- 1,440 units
- Distribution
- US: No distribution OUS: Austria, Belgium, Italy, Spain, Switzerland, United Kingdom
- Product codes and lots
- Lot # HG81GLZ/ GTIN: 20763000649580
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0033-2026 | Class II | Ongoing | Voluntary: Firm initiated |