Z-0033-2026Class II

Medtronic, Simplera Sensor, REF: MMT-5100JD1

Medtronic, Inc. / initiated 2025-05-07 / Open, Classified / root cause: manufacturing process

Medtronic MiniMed, Inc. began a recall of Medtronic, Simplera Sensor, REF: MMT-5100JD1 on . FDA classified it as Class II and gives the reason as "The device cap label may not adhere to the sensor, compromising the sterility of the sensor, and increasing the risk of infection". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0033-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Medtronic, Inc.
Product
Medtronic, Simplera Sensor, REF: MMT-5100JD1
Product code
MDS Sensor, Glucose, Invasive
Reason for recall
The device cap label may not adhere to the sensor, compromising the sterility of the sensor, and increasing the risk of infection.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
On May 7, 2025, Medtronic issued a "Urgent Field Safety Notice" to affected consignees. Medtronic asked consignees to take the following actions: 1. Verify if the sensors you currently have are from the impacted lot. 2. If you have a Simplera" sensor from the impacted lot, do not use this Simplera" sensor and dispose of it according to local regulations. 3. Please contact our helpline/your Medtronic contact at 01923 205167 to request a replacement. 4. Please acknowledge that you have read and understood this notification and have followed the actions listed in this letter, even if your sensor is unaffected, or if you do not have any sensors from the affected lot.
Quantity in commerce
1,440 units
Distribution
US: No distribution OUS: Austria, Belgium, Italy, Spain, Switzerland, United Kingdom
Product codes and lots
Lot # HG81GLZ/ GTIN: 20763000649580

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0033-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0033-2026Class IIOngoingVoluntary: Firm initiated