Z-1925-2025Class II

Brand Name: OBM00002 OBM DAB (Digital Acquisition Box) Product Name: OBM00002 OBM DAB (Digital Acquisition...

Natus Medical Incorporated / initiated 2025-05-07 / Open, Classified / root cause: manufacturing process

Natus Neurology DBA Excel Tech., Ltd. (XLTEK) began a recall of Brand Name: OBM00002 OBM DAB (Digital Acquisition Box) Product Name: OBM00002 OBM DAB (Digital Acquisition Box) Model/Catalog Number: OBM00002 OBM DAB (Digital... on . FDA classified it as Class II and gives the reason as "Electrode placement labels on the Digital Acquisition Box (DAB) were applied backwards. This may ultimately lead to misdiagnosis of seizure location in neonates monitored with this device". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1925-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Natus Medical Incorporated
Product
Brand Name: OBM00002 OBM DAB (Digital Acquisition Box) Product Name: OBM00002 OBM DAB (Digital Acquisition Box) Model/Catalog Number: OBM00002 OBM DAB (Digital Acquisition Box) Software Version: N/A Product Description: OBM00002 OBM DAB (Digital Acquisition Box) Component: No
Product code
OMA Amplitude-Integrated Electroencephalograph
Reason for recall
Electrode placement labels on the Digital Acquisition Box (DAB) were applied backwards. This may ultimately lead to misdiagnosis of seizure location in neonates monitored with this device.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
On May 7, 2025, the firm began notifying customers through "Urgent Medical Device Recall" letters. Customers are instructed to check their Olympic Brainz Monitor stock and determine if they have any devices with the incorrect label orientation. Immediately cease use of incorrectly labeled product and quarantine until you receive return instructions from Natus. Customers with incorrect product labelling will be provided with replacement device(s). If you have any questions regarding this notice, please contact [email protected].
Quantity in commerce
3069
Distribution
Worldwide distribution - US Nationwide including Puerto Rico (PR) and OUS (foreign) countries of: Italy (IT), Estonia (EE), Israel (IL), China (CN), Germany (DE), Saudi Arabia (SA), Switzerland (CH), Chile (CL), Colombia (CO), Netherlands (the) (NL), Japan (JP), United Kingdom of Great Britain and Northern Ireland(the) (GB), Russian Federation (the)(RU), Bolivia (Plurinational State of) (BO), Canada (CA), France (FR), Peru (PE), Guyana (GY), Mexico (MX), Belgium (BE), Thailand (TH), Paraguay (PY), Hong Kong (HK), United Arab Emirates (AE), Austria (AT), India (IN), Indonesia (ID), Kuwait (KW), Qatar (QA), Lithuania (LT), Denmark (DK), Myanmar (MM), Uruguay (UY), Tunisia (TN), Turkey (TR), Korea (the Republic of) (KR), Hungary (HU), Slovenia (SI), Argentina (AR), Norway (NO), Oman (OM), Poland (PL), Finland (FI), Ireland (IE), Australia (AU), Latvia (LV), Ecuador (EC), Singapore (SG), Spain (ES), Serbia (RS), South Africa (ZA), Romania (RO), Algeria (DZ), Portugal (PT), Dominican Republic (the) (DO), Lebanon (LB), Malaysia (MY), Costa Rica (CR), Viet Nam (VN), & Bahrain (BH).
Product codes and lots
UDI-DI 00382830010825 Serial Numbers 001 through OBM00002H3613. Note earlier version serial numbers did not contain the sequence of "OBM"

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1925-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1925-2025Class IIOngoingVoluntary: Firm initiated