ELEOS COLLAR STEM, CEMENTED, FLUTED, 15MM X 120MM. Model Number: HC-15120-03M. Component of Limb Salvage...
Onkos Surgical, Inc. began a recall of ELEOS COLLAR STEM, CEMENTED, FLUTED, 15MM X 120MM. Model Number: HC-15120-03M. Component of Limb Salvage System with BioGrip on . FDA classified it as Class II and gives the reason as "Mislabeling of 13mm Modular Segmental Stem package (HC-13120-03M) which included an implant that was 15mm in diameter (HC-15120-03M) instead of being 13mm as stated on the package. The corresponding 15mm cemented stems...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1818-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Onkos Surgical, Inc.
- Product
- ELEOS COLLAR STEM, CEMENTED, FLUTED, 15MM X 120MM. Model Number: HC-15120-03M. Component of Limb Salvage System with BioGrip.
- Product code
- JDI Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented
- Reason for recall
- Mislabeling of 13mm Modular Segmental Stem package (HC-13120-03M) which included an implant that was 15mm in diameter (HC-15120-03M) instead of being 13mm as stated on the package. The corresponding 15mm cemented stems were mislabeled as size 13mm cemented stems.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling Change Control
- Action
- ONKOS Surgical notified their distributors on about 12/22/2023 via email. They were instructed to segregate and quarantine the product immediately, complete and return the "Product and Distribution Information" form within 2 business days and return any existing implants on hand within 2 business days after signing the acknowledgement. They were also instructed that they will receive replacement implants and to place them in their modular segmental collar stem implant packs.
- Quantity in commerce
- 23 units
- Distribution
- US Nationwide distribution in the states of California, Kentucky, and Wisconsin.
- Product codes and lots
- Model Number: HC-15120-03M. UDI-DI: B278HC1512003M0. Serial Numbers: P230274-1XX (XX represents 23 parts in sequential order from 101-123).
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1817-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1818-2025 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.