Brand Name: Catalyst+ Product Name: C4D software used in conjunction with Catalyst (SP001-0025), Catalyst HD...
C-RAD POSITIONING AB began a recall of Brand Name: Catalyst+ Product Name: C4D software used in conjunction with Catalyst (SP001-0025), Catalyst HD (SP001-0026), Catalyst PT (SP001-0027), Catalyst... on . FDA classified it as Class II and gives the reason as "Software issue with scanning equipment that can results in the filed rotation not being applied correctly". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1858-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- C-Rad Positioning AB
- Product
- Brand Name: Catalyst+ Product Name: C4D software used in conjunction with Catalyst (SP001-0025), Catalyst HD (SP001-0026), Catalyst PT (SP001-0027), Catalyst X4 (SP002-0035), Catalyst+ (SP003-0001), Catalyst+ HD (SP003-0002), Catalyst+ PT (SP003-0008), Catalyst+ X4, (SPOO3-0009) Product Description: Patient positioning device, diagnostic imaging/radiotherapy, laser Component: Not a component
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- Software issue with scanning equipment that can results in the filed rotation not being applied correctly.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- On April 22, 2025, C-RAD Positioning AB issued a Field Safety Notice to affected consignees via E-Mail. C-RAD asked consignees to take the following actions: 1. As an immediate action, the affected users are recommended to: - Do not use Site Groups for treatment plans containing non-coplanar fields. - Ungroup all existing Site Groups. 2. Ensure that all the staff operating the Catalyst family systems and the c4D software application are aware of this information. 3. All customers are requested to reply to C-RAD using the attached FSN Customer Reply Form .
- Quantity in commerce
- 44 units
- Distribution
- Worldwide - US Nationwide distribution in the states of AK, FL, WA, VA, WA, NC, IL, AR, NC and the countries of Germany, Italy, Uzbekistan, South Korea, Vietnam, Austria, Australia, China, Estonia, Belgium, Uzbekistan.
- Product codes and lots
- All Catalyst products using software Version: c4D 6.5.1 SP2 Catalyst -SP001-0025/UDI: 07350005280024; Catalyst HD- SP001-0026/UDI: 07350005280024; Catalyst PT- SP001-0027/UDI: 07350005280024; Catalyst X4- SP002-0035/UDI: 07350005280024; Catalyst+ - REF: SP003-0001/UDI: 17350005280762 Catalyst+ HD- SP003-0002/UDI: 17350005280779 Catalyst+ PT- SP003-0008/UDI: 17350005280793 Catalyst+ X4 - SP003-0009/UDI: 17350005280809
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1858-2025 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY