Z-1861-2025Class II
Plus Initial Drill, Article Number: 103.170
Straumann USA, LLC / initiated 2025-04-14 / Open, Classified
Straumann USA LLC began a recall of Plus Initial Drill, Article Number: 103.170 on . FDA classified it as Class II and gives the reason as "The referred batch was produced without the laser engraving of 7 mm". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-1861-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Straumann USA, LLC
- Product
- Plus Initial Drill, Article Number: 103.170;
- Product code
- DZA Drill, Dental, Intraoral
- Reason for recall
- The referred batch was produced without the laser engraving of 7 mm.
- FDA root cause
- Under Investigation by firm
- Action
- On April 14, 2025 customers were notified via phone with a follow-up letter sent on April 28, 2024. Both contacts included the following: Action to be taken: 1. If article 103.170 from lot LNWT8 is still in your inventory, then stop use/distribution of the product immediately and quarantine/segregate physically. 2. If the product was used and no complications or problems were identified, of the patient did not present symptoms there is no need for any additional action with the patient. 3. If article 103.170 from Lot LNWT8 is still in your inventory, return it for a credit or replacement as indicated on the Customer Confirmation Form. 4. In the product is not found in your stock, indicate this on the enclosed Customer Confirmation Form. 5 In all cases, complete and return the enclosed Customer Confirmation Form with the affected product using the return UPS label provided to: Straumann USA, LLC; 60 Minuteman Road, Andover, MA 01810. transmission of the Field Safety Notice This notice needs to be passed on to all those who need to be aware within your organization or to any organization where the affected devices have been transferred. The FDA has been informed about this Field Safety Corrective Action. We apologize for any inconvenience that this may cause.
- Quantity in commerce
- 288 units ($& US, 241 OUS)
- Distribution
- Worldwide - US Nationwide distribution in the states of AZ, CA, FL, GA, IL, IN, MN, NC, NV, NY, OH, OK, SC, TX, VA, WI and the countries of Brazil, Germany, India, Indonesia, Portugal, Romania, Spain, Switzerland, Ukraine.
- Product codes and lots
- Article Number: 103.170; UDI-DI: (01)7898237561967(11)250204(17)400204(10)LNWT8; Lot Number: LNWT8