Medline Kits containing Codman Disposable Perforator 14 mm: 1. CRANI ADD A PACK, SKU DYNJ30926O 2...
MEDLINE INDUSTRIES, LP - Northfield began a recall of Medline Kits containing Codman Disposable Perforator 14 mm: 1. CRANI ADD A PACK, SKU DYNJ30926O 2. CRANIOTOMY, SKU CDS983611N 3. CRANIOTOMY CDS, SKU CDS983467J... on . FDA classified it as Class I and gives the reason as "Affected kits contain recalled Integra Lifesciences Codman Disposable Perforators due to inadequate welds on specific lots. Perforators with inadequate welds may break or separate". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1809-2025
- Classification
- Class I
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Product
- Medline Kits containing Codman Disposable Perforator 14 mm: 1. CRANI ADD A PACK, SKU DYNJ30926O 2. CRANIOTOMY, SKU CDS983611N 3. CRANIOTOMY CDS, SKU CDS983467J 4. CRANIOTOMY PACK, SKU DYNJ85807A, DYNJ85927 5. MAJOR CRANIOTOMY PACK, SKU DYNJ82007A
- Product code
- LRO General Surgery Tray
- Reason for recall
- Affected kits contain recalled Integra Lifesciences Codman Disposable Perforators due to inadequate welds on specific lots. Perforators with inadequate welds may break or separate.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- On April 22, 2025, the firm notified customers of the recall through letters titled Immediate Action Required Voluntary Medical Device Recall. Customers were instructed to identify and quarantine all affected product on hand. Customers are to submit a response form to the recalling firm. Upon receipt of the response firm, Medline will provide the customer with over-labels to place on affected inventory, which will include instructions for staff to remove and discard the affected component prior to using the kit. Distributors should notify downstream customers/accounts of this recall. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].
- Quantity in commerce
- 1853 total
- Distribution
- US distribution to CA, FL, IL, IN, KY, MA, MD, MN, NY, TX, VT, WA and WI. No OUS distribution.
- Product codes and lots
- 1. CRANI ADD A PACK, SKU DYNJ30926O, UDI-DI 10195327539153 (EA), 40195327539154 (CASE) KIT LOTS 24AMF811 2. CRANIOTOMY, SKU CDS983611N, UDI-DI 10195327466626 (EA), 40195327466627 (CASE) KIT LOTS 23KBA720 23LBI611 23LBJ586 3. CRANIOTOMY CDS, SKU CDS983467J, UDI-DI 10195327561116 (EA), 40195327561117 (CASE) KIT LOTS 24ADA818 24ADC216 24BDA929 24DDB447 24EDA275 24EDA751 24GDB213 24IDA327 24IDB043 4. CRANIOTOMY PACK, SKU DYNJ85807A, UDI-DI 10195327637392 (EA), 40195327637393 (CASE) KIT LOTS 24CBB843 24EBC889 24GBY029 ; SKU DYNJ85927, UDI-DI 10195327575571 (EA), 40195327575572 (CASE) KIT LOTS 24BMD310 24EMI836 5. MAJOR CRANIOTOMY PACK, SKU DYNJ82007A UDI-DI 10195327635466 (EA), 40195327635467 (CASE) KIT LOTS 24CMA496 24HMI081 24HMI978
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1809-2025 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-1810-2025 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-1811-2025 | Class I | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY