Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1909-2025Class II

Pre-filled Water Syringe, Model Numbers SW-010-100 and SW-010

Healthmark Industries Co., Inc. / initiated 2025-05-02 / Open, Classified

Healthmark Industries Co., Inc. began a recall of Pre-filled Water Syringe, Model Numbers SW-010-100 and SW-010 on . FDA classified it as Class II and gives the reason as "During endoscope surveillance monitoring, a user facility identified contamination from duodenoscope samples that were linked to the Pre-filled Water Syringes as part of their investigation. Healthmark Industries...". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-1909-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Healthmark Industries Co., Inc.
Product
Pre-filled Water Syringe, Model Numbers SW-010-100 and SW-010
Product code
MNL Accessories, Cleaning Brushes, For Endoscope
Reason for recall
During endoscope surveillance monitoring, a user facility identified contamination from duodenoscope samples that were linked to the Pre-filled Water Syringes as part of their investigation. Healthmark Industries conducted its own investigation and concluded that the Pre-filled Water Syringe had been improperly used outside of the intended purpose.
FDA root cause
Under Investigation by firm
Action
An "URGENT PRODUCT RECALL" notification letter was sent to customers beginning 5/2/25. Recommended Mitigations/ User Recommendations Please discontinue use of the Pre-Filled Water Syringes containing non-sterile water, which can be identified by a red cap. Any unused product can be returned for credit or replacement. Customer are requested to take the following actions 1. Ensure all affected product in your facility s inventory is no longer in use and has been properly quarantined. 2. Please ensure this message is forwarded to any individuals that need notification within your organization or any organization where the affected products have been transferred. 3. Contact Customer Service via email at [email protected] or 800-521-6224 to return your unused affected product or if you have any questions. 4. Complete and return the response form Getinge by e-mailing a copy to [email protected].
Quantity in commerce
250000 units
Distribution
US Nationwide and Internationally to countries of: Canada, Malaysia.
Product codes and lots
Lot Numbers: 200001 200002 200003 200004 200005 200006 200007 200008 200009 200010 200011 200012 200013

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1909-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1909-2025Class IIOngoingVoluntary: Firm initiated
Z-1910-2025Class IIOngoingVoluntary: Firm initiated