LOWER EXTREMITY, Model: NCEX20I; LOWER EXTREMITY, Model: NCEX20J; TOTAL KNEE PACK, Model: NCTK31K; TOTAL KNEE...
American Contract Systems, Inc. began a recall of LOWER EXTREMITY, Model: NCEX20I; LOWER EXTREMITY, Model: NCEX20J; TOTAL KNEE PACK, Model: NCTK31K; TOTAL KNEE PACK, Model: NCTK31L All convenience kits contain... on . FDA classified it as Class II and gives the reason as "Test results for cast padding component included in procedure trays shows Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals exceeded residual limits, which may lead to a possible loss or lack of functionality that...". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-1916-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- American Contract Systems, Inc.
- Product
- LOWER EXTREMITY, Model: NCEX20I; LOWER EXTREMITY, Model: NCEX20J; TOTAL KNEE PACK, Model: NCTK31K; TOTAL KNEE PACK, Model: NCTK31L All convenience kits contain Specialist Blend Cast Padding, Item: QTX30-322NS
- Product code
- OJH Orthopedic Tray
- Reason for recall
- Test results for cast padding component included in procedure trays shows Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals exceeded residual limits, which may lead to a possible loss or lack of functionality that may lead to a delay or prolonged treatment/procedure/therapy.
- FDA root cause
- Finished device change control
- Action
- On 4/21/2025, correction notices were emailed to the customer who was asked to do the following: 1) Discontinue use of the Cast Padding component. All other components within the kit can be used. 2) Add warning labels to affected kits in Table 2 to ensure the component is discarded. A labeling template is enclosed for printing labels which can be affixed on all affected kits in inventory. This label should be applied to each affected Procedure tray and product case in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling. - Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. - Distributors: If any of the affected procedure trays identified above were further distributed, please ensure all end-users are appropriately notified of this product correction and maintain records of effectiveness. You are also responsible for the reconciliation of these customers. Please do not have your customers respond directly to ACS. Complete and return the response form via email to [email protected] Report any quality problems or adverse events associated with the products listed in this notification to [email protected]
- Quantity in commerce
- 850 Kits
- Distribution
- US Nationwide distribution in the state of CO.
- Product codes and lots
- Model/UDI-DI/Lot(Expiration): NCEX20I/00191072222398/9146304(2025-12-03), 9207685(2026-01-08); NCEX20J/00191072234582/9222096(2026-01-30), 9222096(2026-01-30); NCTK31K/00191072222480/9086225(2025-11-01), 9086225(2025-11-01); NCTK31L/00191072234872/9219596(2026-01-15), 9219596(2026-01-15)
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1916-2025 | Class II | Ongoing | Voluntary: Firm initiated |