Z-2151-2025Class II

Covidien Nellcor Bedside SpO2 Patient Monitoring System: REF: PM100N, 10005941, DLPM100N, DL10005941...

Medtronic, Inc. / initiated 2025-06-23 / Open, Classified / root cause: labeling and UDI

Covidien began a recall of Covidien Nellcor Bedside SpO2 Patient Monitoring System: REF: PM100N, 10005941, DLPM100N, DL10005941, DSPM100N, 10005941-SG, DS10005941, 10005941J, 10005941JP... on . FDA classified it as Class II and gives the reason as "Bedside SpO2 Patient Monitoring System alarms not heard/recognized, resulting in: treatment delay, lack of low oxygen saturation response, respiratory failure, arrhythmia. Highlighting operator's manual: 1)For patient...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-2151-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Medtronic, Inc.
Product
Covidien Nellcor Bedside SpO2 Patient Monitoring System: REF: PM100N, 10005941, DLPM100N, DL10005941, DSPM100N, 10005941-SG, DS10005941, 10005941J, 10005941JP, 10005941JPN
Product code
DQA Oximeter
Reason for recall
Bedside SpO2 Patient Monitoring System alarms not heard/recognized, resulting in: treatment delay, lack of low oxygen saturation response, respiratory failure, arrhythmia. Highlighting operator's manual: 1)For patient Safety don't silence/decrease alarm volume, 2)Sensor off alarm (no perfusion detected) is low priority alarm, 3)Set monitors to Homecare Mode outside of hospital/professional setting
Root cause tag
labeling and UDI
FDA root cause
Labeling design
Action
On 6/23/2025 correction notices were distributed to customers who where informed: Per Operator s Manual (http://manuals.medtronic.com): 1) Device alarm volume needs to be set to a level appropriate to be heard throughout the care environment at all times of day or night. Do not silence or decrease the volume of the audible alarm if patient safety could be compromised. 2) If the sensor comes off the patient, or no perfusion is detected, the device will notify 'sensor off' (low priority alarm) every 16 seconds and display a steady yellow visual indicator. If a low priority alarm immediately follows a higher priority alarm, a user may not recognize the higher priority clinical condition. 3) Monitors should be set to Standard Mode when in use in a hospital or hospital-type environment by trained medical personnel and set to Homecare Mode when a lay person will be using the monitor outside of a hospital or other professional care setting. With the monitor not in Homecare Mode, alarms could be silenced/muted and alarm limits altered In Clinical Facility Environments be sure the PM100N model is set to Standard Mode for general use and only set to Sleep Study Mode when conducting a sleep study. Device should be set to Homecare Mode prior to use. Firm's Technical Service at 800-255-6774 (option 1, then 2) can provide a four-digit password required to set a monitor to Homecare Mode. Provide the Home Use Guide to home caregivers. Clinicians should communicate the recall information to home use patients and caregivers. Pass on/post this notice for all those who need to be aware within your organization and to any organization where the product has been transferred/distributed. Complete and return the Customer Confirmation Form to [email protected] Firm will provide additional information when finalized. Questions - Contact firm's Customer Service at 800-962-9888, Option 2.
Quantity in commerce
264,790
Distribution
Worldwide - US Nationwide distribution including in the states of MS, TX, FL, SC, VA, AZ, PA, MN, OH, CA, MI, CO, KS, IN, IL, TN, MO, IA, OK, WI, AK, NY, MD, AL, WV, GA, NC, AR, WA, SD, KY, DE, NJ, UT, NV, MA, MT, GU, NE, LA, HI, OR, RI, NH, ND, NM, ME, CT, DC, WY and the countries of Algeria, Andorra, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Bermuda, Bhutan, Bolivia, Bosnia And Herzegovina, Botswana, Brazil, Brunei Darussalam, Bulgaria, Cambodia, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, French Guiana, Germany, Ghana, Greece, Guadeloupe, Guatemala, Hong Kong, Hungary, India, Indonesia, Iran, Islamic Republic Of, Ireland, Isle of Man, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Maldives, Malta, Mauritius, Mexico, Montenegro, Morocco, Myanmar, Namibia, Netherlands, New Zealand, North Macedonia, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Vietnam, Zimbabwe.
Product codes and lots
REF/UI-DI/Bundle Code/IFU Part Number: PM100N/10884521196728/NELL-HOMECARE, BNPULSOKSYMETR, PM100N, PM100NKIT, PM100NKITCHDE, PM100NKITCHFR, PM100NKITDA, PM100NKITDE-2, PM100NKITES, PM100NKITFI, PM100NKITFR, PM100NKIT-FR, PM100NKITIT, PM100NKIT-NL, PM100NKITNO, PM100NKITPT, PM100NKITSV, PM100NKITUK, PM100N-10, PM100N-2XMAXN, PM100N-2XMAXN-CC, PM100N-3DYS-CC, PM100N-DYS, PM100N-HC, PM100N-MAXN, PM100N-MAXN-CC, PM100N-OXIAN-CC/PT00156589 - AR - Arabic, PT00156609 - AR - Arabic, PT00156588 - BG - Bulgarian, PT00156610 - BG - Bulgarian, PT00156586 - CS - Czech, PT00156612 - CS - Czech, PT00156585 - DA - Danish, PT00156613 - DA - Danish, PT00156578 - DE - German, PT00156625 - DE - German, PT00156577 - EL - Greek, PT00156626 - EL - Greek, PT00156509 - EN - English, PT00156328 - EN - English, PT00156563 - ES - Spanish, PT00156639 - ES - Spanish, PT00156583 - ET - Estonian, PT00156615 - ET - Estonian, PT00156582 - FI - Finnish, PT00156616 - FI - Finnish, PT00156581 - FR - French, PT00156617 - FR - French, PT00156587 - HR - Croatian , PT00156611 - HR - Croatian , PT00156576 - HU - Hungarian, PT00156627 - HU - Hungarian, PT00156575 - IT - Italian, PT00156628 - IT - Italian, PT00156580 - KA - Georgian, PT00156618 - KA - Georgian, PT00156591 - KO - Korean, PT00156642 - KO - Korean, PT00156573 - LT - Lithuanian, PT00156630 - LT - Lithuanian, PT00156574 - LV - Latvian, PT00156629 - LV - Latvian, PT00156584 - NL - Dutch, PT00156614 - NL - Dutch, PT00156572 - NO - Norwegian, PT00156631 - NO - Norwegian, PT00156571 - PL - Polish, PT00156632 - PL - Polish, PT00156570 - PT-BR - Brazilian Portuguese, PT00156633 - PT-BR - Brazilian Portuguese, PT00156569 - RO - Romanian, PT00156634 - RO - Romanian, PT00156567 - RU - Russian, PT00156635 - RU - Russian, PT00156565 - SK - Slovak, PT00156637 - SK - Slovak, PT00156564 - SL - Slovenian, PT00156638 - SL - Slovenian, PT00156566 - SR - Serbian, PT00156636 - SR - Serbian, PT00156562 - SV Swedish, PT00156640 -SV - Swedish, PT00156561 - TR - Turkish, PT00156641 - TR - Turkish, PT00156590 - ZH - SimplifIed Chinese, PT00156643 - ZH - SimplifIed Chinese; 10005941/10884521163454/PM100N-10, PM100N-2XMAXN, PM100N-2XMAXN-CC, PM100N-3DYS-CC, PM100N-DYS, PM100N-HC, PM100N-MAXN, PM100N-MAXN-CC, PM100N-OXIAN-CC, BEDSIDESP2KIT2, NELLSPO2, PM100N-HC, NELL-HOMECARE, ROEBEDSIDEKIT, BEDSIDESP2PT2, BEDSIDESP2NO2, BEDSIDESP2PL2, BEDSIDESP2NL2, BEDSIDESP2KITNM, BEDSIDESP2K2UK, BEDSIDESP2K2IT, BEDSIDESP2K2FR, BEDSIDESP2K2-2, BEDSIDEKITISR/PT00156645 - AR - Arabic, PT00156647 - BG - Bulgarian, PT00156649 - CS - Czech, PT00156650 - DA - Danish, PT00156656 - DE - German, PT00156657 - EL - Greek, PT00156324 - EN - English, PT00156673 - ES - Spanish, PT00156652 - ET - Estonian, PT00156653 - FI - Finnish, PT00156654 - FR - French, PT00156648 - HR - Croatian, PT00156658 - HU - Hungarian, PT00156659 - IT - Italian, PT00156655 - KA - Georgian, PT00156677 - KO - Korean, PT00156661 - LT - Lithuanian, PT00156660 - LV - Latvian, PT00156651 - NL - Dutch, PT00156662 - NO - Norwegian, PT00156664 - PL - Polish, PT00156666 PT-BR - Brazilian Portuguese, PT00156667 - RO - Romanian, PT00156668 - RU - Russian, PT00156671 - SK - Slovak, PT00156672 - SL - Slovenian, PT00156669 - SR - Serbian, PT00156674 - SV - Swedish, PT00156675 - TR - Turkish, PT00156676 - ZH - Simplified Chinese, NELL-SPO2-AD3+ - Canadian, NELL-SPO2-AD3 - Canadian; 10005941/A8845211634501; DLPM100N/10884521527607; DL10005941/10884521173293; DSPM100N/DSPM100N; 10005941-SG/10884521171534/BNINHLLCODE, BNINNEL100A, BNINNEL100AS, BNINNEL100AVJ, BNINNEL100AYS, BNINNELDYS, BNHKLIBRANEOSET, BNHKLIBRASET; DS10005941/DS10005941; 10005941J/10005941JP; 10005941JP/884521512019; 10005941JPN/884521188689. All serial numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2151-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2151-2025Class IIOngoingVoluntary: Firm initiated