Model: Description/REF: MNA-420: MNA-420 ENC -H/K9303320, MNA-420 DEC -H V2/K9303331; MNA-440: MNA-440 ENC...
Barco N.V. began a recall of Model: Description/REF: MNA-420: MNA-420 ENC -H/K9303320, MNA-420 DEC -H V2/K9303331; MNA-440: MNA-440 ENC -H/K9303300, MNA-440 DEC -H V2/K9303311; MNA-620... on . FDA classified it as Class II and gives the reason as "Cathode ray tube display system encoders and decoders have same MAC address; new system setup with two or more affected units can lead to systems not functioning reliably:audio, video and/or control signals might be...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1994-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Barco N.V.
- Product
- Model: Description/REF: MNA-420: MNA-420 ENC -H/K9303320, MNA-420 DEC -H V2/K9303331; MNA-440: MNA-440 ENC -H/K9303300, MNA-440 DEC -H V2/K9303311; MNA-620: MNA-620 DEC -H/K9306210, MNA-620 ENC -H/K9306220; and MNA-640: MNA-640 DEC -H/K9306410, MNA-640 ENC -H/K9306420; components of the NexxisOR system
- Product code
- DXJ Display, Cathode-Ray Tube, Medical
- Reason for recall
- Cathode ray tube display system encoders and decoders have same MAC address; new system setup with two or more affected units can lead to systems not functioning reliably:audio, video and/or control signals might be interrupted or fall away, displays might switch to failover and back multiple times, devices/sources/sinks might not be discovered/available, might lead to delayed/prolonged procedures
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- On 5/9/2025, recall notices started being distributed to customers informing them of the following: 1) Assess & locate the non-installed affected units and contain them from further distribution. B) Assess & locate affected units that are installed at your customer side. Please make your customers aware of this notification. C) The impacted units will need to be sent back. Further information on the swaps will be provided in the coming days. D) Fill in and return the Acknowledgement Letter. If you don t have the Key Account Manager s contact details, or need other support, you can contact the firm 24/7 via their Support web page: https://www.barco.com/en/support
- Quantity in commerce
- 221
- Distribution
- US Nationwide distribution in the states of TX, OH.
- Product codes and lots
- REF/UDI-DI/Serial Number: K9303320/05415334002364/2531531598, 2531531604, 2531534397, 2531534420, 2531534430, 2531534466, 2531534905, 2531534948, 2531534956, 2531536564, 2531554560, 2531572616, 2531572628, 2531572634, 2531591962, 2531591973, 2531591992, 2531592064, 2531592096, 2531593814, 2531593818, 2531593823, 2531597597, 2531610546, 2531610547, 2531610572, 2531610578, 2531610580, 2531612404, 2531612420, 2531612424, 2531612430, 2531612431, 2531612432, 2531612433, 2531612434, 2531612436, 2531612442, 2531623109, 2531623114, 2531623131, 2531623133, 2531623141, 2531635578, 2531635582, 2531639224, 2531639261, 2531639263, 2531639353, 2531640594, 2531640597, 2531640598, 2531640603, 2531640626, 2531640634, 2531540640, 2531640646, 2531640658, 2531640660; K9303331/05415334026827/2531533275, 2531533289, 2531533292, 2531533293, 2531533304, 2531533310, 2531533312, 2531533314, 2531533316, 2531533353, 2531533356, 2531534284, 2531534292, 2531534362, 2531537718, 2531540799, 2531569738, 2531582287, 2531582290, 2531582311, 2531582315, 2531582378, 2531582750, 2531582755, 2531582790, 2531582791, 2531582826, 2531582827, 2531582829, 2531584018, 2531584044, 2531537696, 2531537726; K9303300/05415334002180/2531537815, 2531557845, 2531557866, 2531558588, 2531558727, 2531558738, 2531558753, 2531558754, 2531558759, 2531561761, 2531561785, 2531572541, 2531572557, 2531572568, 2531572576, 2531572847, 2531572862, 2531572872, 2531633606, 2531633662; K9303311/05415334026810/2531537674, 2531572141, 2531580810, 2531580935, 2531582319, 2531582331, 2531582341, 2531588386, 2531588401, 2531588404, 2531588456, 2531591560, 2531592036, 2531592037, 2531592046, 2531593872, 2531594457, 2531594475, 2531594481, 2531594503, 2531594515, 2531597664, 2531597713, 2531597717, 2531597726, 2531597795, 2531597806, 2531597809, 2531597810, 2531597820, 2531599836, 2531605356, 2531605390; K9306210/2531542260, 2531542263, 2531542270, 2531542292, 2531543553, 2531543568, 2531543588, 2531582676, 2531582691, 2531588558, 2531588562, 2531593741, 2531593749, 2531593752, 2531593763, 2531593764, 2531593768, 2531593777, 2531593784, 2531595829, 2531595831, 2531601786; K9306220/ 2531540107, 2531542441, 2531542470, 2531543538, 2531543549, 2531543611, 2531543613, 2531543640, 2531563132, 2531572740, 2531572759, 2531572770, 2531572814, 2531572827, 2531575652, 2531594331, 2531594332, 2531594336, 2531594348, 2531594350; K9306410/2531572228, 2531582713, 2531582724, 2531582725, 2531584168, 2531584176, 2531584191, 2531588591, 2531588593, 2531588598, 2531588616, 2531588626, 2531604519, 2531604535, 2531604544, 2531604545, 2531604549, 2531604550, 2531604551, 2531604593, 2531604637, 2531604649, 2531604658; K9306420/ 2531537844, 2531537870, 2531537889, 2531537901, 2531537902, 2531556037, 2531556043, 2531572419, 2531572433, 2531572440, 2531595759
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1994-2025 | Class II | Ongoing | Voluntary: Firm initiated |