Brand Name: Pluslife SARS-CoV-2 Card Product Name: SARS-CoV-2 Nucleic Acid Testing Card Model/Catalog Number...
Guangzhou Pluslife Biotech Co., Ltd. began a recall of Brand Name: Pluslife SARS-CoV-2 Card Product Name: SARS-CoV-2 Nucleic Acid Testing Card Model/Catalog Number: RM1010202 Software Version: N/A Product... on . FDA classified it as Class II and gives the reason as "Devices were distributed in the U.S. market without obtaining the required FDA premarket authorization (e.g., 510(k) clearance or PMA)". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-1979-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Guangzhou Pluslife Biotech Co., Ltd.
- Product
- Brand Name: Pluslife SARS-CoV-2 Card Product Name: SARS-CoV-2 Nucleic Acid Testing Card Model/Catalog Number: RM1010202 Software Version: N/A Product Description: 1. Product Overview Product Name: SARS-CoV-2 Nucleic Acid Testing Card Trade Name: Pluslife SARS-CoV-2 Card 2.Product composition:The kits consists of SARS-CoV-2 Reaction Card (10 pcs), Nucleic Acid ReleasingAgent 01 (10 pcs), Disposable SamplingSwab (10 pcs), and Waste Bag (10 pcs). Component: NO
- Product code
- QKP Coronavirus Antigen Detection Test System.
- Reason for recall
- Devices were distributed in the U.S. market without obtaining the required FDA premarket authorization (e.g., 510(k) clearance or PMA).
- FDA root cause
- No Marketing Application
- Action
- On January 24, 2025, FORMAL NOTICE REGARDING DISTRIBUTION RESTRICTIONS IN THE U.S. MARKET letters were emailed to distributors. This communication serves to formally notify you that, effective immediately, all authorized distributors of Guangzhou Pluslife Biotech Co., Ltd. are strictly prohibited from marketing, selling, or distributing any Pluslife products to the United States market. In May 2025 an "URGENT MEDICAL DEVICE RECALL NOTIFICATION" letters dated May 30, 2025, were sent to the distributor to be sent to consignees. Actions to be taken by the Customer/User: Consumers who still have unused tests manufactured by Pluslife should stop using the device immediately and discard any remaining stock in the trash. - Notice to Distributor: Immediately cease U.S. sales of unauthorized Pluslife devices Issue recall notices to all affected customers in the U.S cooperate fully with the FDA and Pluslife to complete the recall - Company official statement Statement on Prohibition of Unauthorized Sales to the U.S. Market (Published on the Pluslife official website). - Signing the distribution agreement: Prohibiting distributors from selling the unapproved Pluslife products by the FDA in U.S. For assistance, contact: Manufacturer by Pluslife: Marketing Team [email protected] Mon-Fri, 9:00 AM to 6:00 PM CST Distributor by Altruan : After-Sales Team [email protected] Mon-Fri, 9:00 AM to 5:00 PM CET
- Quantity in commerce
- 1850 boxes
- Distribution
- Worldwide distribution: US (nationwide) and OUS (foreign) to countries of: Hong Kong, Australia, Singapore, Palestine, Russia, Ecuador, Papua New Guinea, Indonesia, South Africa, Zimbabwe, Brazil, Tunisia, Samoa, Algeria, Romania, Czech Republic, Thailand, South Korea, Malaysia, Fiji, Nepal, Peru, Mexico, Bahamas, Chile, India, Germany, France, Portugal, Vietnam, Spain, Poland, Latin American Region, Italy, Cameroon, Slovakia, Greece, Hungary Bahrain, Estonia, and Ukraine.
- Product codes and lots
- Model/Catalog Number: RM1010202; Lot numbers: All lots of the COVID-19 test kits in the U.S.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1978-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1979-2025 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.