SwabCap Disinfecting Cap for Needlefree Connectors: SwabSleeve, REF: SCXT3-5-2000; SwabCap, 200-ct. Box, REF...
ICU Medical, Inc. began a recall of SwabCap Disinfecting Cap for Needlefree Connectors: SwabSleeve, REF: SCXT3-5-2000; SwabCap, 200-ct. Box, REF: SCXT3-2000 on . FDA classified it as Class II and gives the reason as "Due to a manufacturing issue, disinfecting cap for needle-free connectors may have an incomplete seal between the foil lid and plastic container, which may result in isopropyl alcohol evaporation from the sponge, which...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2116-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Abbott Medical
- Product
- SwabCap Disinfecting Cap for Needlefree Connectors: SwabSleeve, REF: SCXT3-5-2000; SwabCap, 200-ct. Box, REF: SCXT3-2000
- Product code
- QBP Cap, Device Disinfectant
- Reason for recall
- Due to a manufacturing issue, disinfecting cap for needle-free connectors may have an incomplete seal between the foil lid and plastic container, which may result in isopropyl alcohol evaporation from the sponge, which may result in inadequate disinfection.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- On 6/3/2025, correction notices were mailed and emailed to customers who were asked to do the following: 1) Inform all potential users of the product in your organization of this notification. 2) Discard the affected devices following your institution's process for discarding product. If discarding the device is not immediately possible at your facility, the device should be quarantined until disposal is possible. 3) Complete and return the Customer Response Form to [email protected] 5. DISTRIBUTORS: If you have distributed affected products to your customers, please immediately forward this notice to them. Request that they complete the response form and return it to [email protected]. For further inquiries, please contact firm at the following: Global Complaint Management, [email protected],1-(866)-216-8806; Customer Service, [email protected], 1-(800)-258-5361; Field Action Processing, [email protected], 1-(866)-328-0114
- Quantity in commerce
- 1,445,600
- Distribution
- Worldwide - US Nationwide distribution including in the states of TX, SD, ID, LA, IN, CO, OK, MO, MN, OR, AZ, NC, TN, VA, CA, IL, FL, MA, OH, MD, PA, NJ, NY, WA, GA, UT, MI, MS, AR, AL, NV, KS, IA, WV, HI, KY, NM and the country of CA.
- Product codes and lots
- REF/UDI-DI/Lot(Expiration): SCXT3-5-2000/10887709097107/14258739(1-Dec-26), 14258743(1-Feb-26), 14258744(1-Dec-26), 14264200(1-Dec-26), 14272241(1-Dec-26), 14272248(1-Dec-26), 14287456(1-Dec-26), 14287459(1-Dec-26), 14287480(1-Dec-26), 14287482(1-Dec-26), 14287485(1-Dec-26), 14294832(1-Dec-26), 14272254(1-Dec-26), 14310313(1-Dec-26), 14310425(1-Dec-26), 14323354(1-Dec-26), 14287475(1-Dec-26), 14294840(1-Dec-26), 14321645(1-Jan-27), 14321652(1-Jan-27), 14331169(1-Jan-27), 14332773(1-Jan-27), 14341732(1-Oct-26), 14294825(1-Jan-27), 14303209(1-Jan-27), 14310308(1-Jan-27), 14355404(1-Jan-27), 14355409(1-Jan-27) SCXT3-2000/20887709079162/14258214(1-Jan-27), 14258206(1-Jan-27), 14258208(1-Feb-27).
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2116-2025 | Class II | Ongoing | Voluntary: Firm initiated |