LIFEPEARL Drug Elutable Microspheres: 200 +/- 50 micrometers, REF: 8LP2S200; 400 +/- 50 micrometers, REF...
MICROVENTION INC. began a recall of LIFEPEARL Drug Elutable Microspheres: 200 +/- 50 micrometers, REF: 8LP2S200; 400 +/- 50 micrometers, REF: 8LP2S400 on . FDA classified it as Class II and gives the reason as "Drug Elutable Microspheres have a smaller actual average diameter that is not within specification, which may lead to inability to reach the desired treatment location (non-targeted embolization), additional...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2188-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Product
- LIFEPEARL Drug Elutable Microspheres: 200 +/- 50 micrometers, REF: 8LP2S200; 400 +/- 50 micrometers, REF: 8LP2S400
- Product code
- KRD Device, Vascular, For Promoting Embolization
- Reason for recall
- Drug Elutable Microspheres have a smaller actual average diameter that is not within specification, which may lead to inability to reach the desired treatment location (non-targeted embolization), additional procedure/treatment required, blockage other than target vessel, increased procedure time, incomplete embolization, and/or inability to treat the patient.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On 5/29/2025, recall notices were mailed and emailed to customers who were asked to do the following: 1) Inform all individuals within your organization of the recall and forward the recall notice to any organizations that may have received the affected products. 2) Complete and return the customer acknowledgement form via email: [email protected] 3) If your institution has affected inventory call firm's Customer Care at 800-888-3786 or email [email protected] Please direct questions to the firm at [email protected]
- Quantity in commerce
- 1451
- Distribution
- OUS: Italy, Spain, Portugal, Poland, Belgium, Germany, Denmark, France, Kazakhstan, Slovakia, Norway, Greece, Serbia, Turkey, Czech Republic, United Kingdom, Netherlands, Russian Federation, Kuwait, Republic of North Macedonia, Switzerland, Saudi Arabia, Georgia, Lithuania, Romania, Croatia, Chile, Hong Kong , India, Mexico, Taiwan, Vietnam
- Product codes and lots
- REF/GTIN/Lot: 8LP2S20/(01)00812636020105/0000177644, 0000180340, 0000223990, 0000235256, 0000254143, 0000254142, 0000272315, 0000272316, 0000281719, 0000294361, 0000295741, 0000296446, 0000751662, 0000751661, 0001026153; 8LP2S400/(01)00812636020112/0000199869, 0000225205, 0000708506, 0000863001, 0001026156
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2188-2025 | Class II | Ongoing | Voluntary: Firm initiated |