Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2008-2025Class II

Brand Name: ALTIVATE REVERSE Product Name: ALTIVATE REVERSE TORX PERIPHERAL SCREW, 26mm Model/Catalog Number...

Encore Medical, LP / initiated 2025-05-28 / Open, Classified / root cause: manufacturing process

Encore Medical, LP began a recall of Brand Name: ALTIVATE REVERSE Product Name: ALTIVATE REVERSE TORX PERIPHERAL SCREW, 26mm Model/Catalog Number: 506-04-126 Software Version: NA Product... on . FDA classified it as Class II and gives the reason as "Package labeling for Altivate Reverse Torx Peripheral screw contains a different size then actual included product". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2008-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Encore Medical, LP
Product
Brand Name: ALTIVATE REVERSE Product Name: ALTIVATE REVERSE TORX PERIPHERAL SCREW, 26mm Model/Catalog Number: 506-04-126 Software Version: NA Product Description: Encore Medical L.P. shoulder devices are intended for treatment of patients who are candidates for shoulder arthroplasty per the Indication for use. While total shoulder replacements are not intended to withstand activity levels and loads of normal healthy bone, they are a means of restoring mobility and reducing pain for many patients. Component: NA
Product code
PHX Shoulder Prosthesis, Reverse Configuration
Reason for recall
Package labeling for Altivate Reverse Torx Peripheral screw contains a different size then actual included product.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
On May 28, 2025, Enovis (Encore Medical, L.P.) issued a "Urgent: Medical Device Recall" Notification via: E-Mail. Envois is asking consignees to take the following actions: 1. Review your stock for the products and lot numbers for the items to be returned. 2. This notice needs to be passed on to all those within your organization or to any organization where the product has been transferred. 3. Complete acknowledgement and response form and contact Customer Service at 1-800-456-8696 or through your regional email address, to place a replacement request. You will be provided with an RMA number and shipping label to return the affected product. 4. Return all devices (quantities listed above) using the RMA number and shipping label provided by Customer Service.
Quantity in commerce
400 units
Distribution
At time of quarantine: RFK Medical Inc. Virginia (VA) CE102 506-04-122 5035A1018 ALTIVATE REVERSE TORX PERIPHERAL SCREW, 22mm 2 506-04-126 5037A1017 ALTIVATE REVERSE TORX PERIPHERAL SCREW, 26mm 1 Evolution Medical North Dakota (ND), Minnesota (MN), Colorado (CO), Utah (UT) CE106 506-04-126 5037A1017 ALTIVATE REVERSE TORX PERIPHERAL SCREW, 26mm 1 506-04-130 5039A1021 ALTIVATE REVERSE TORX PERIPHERAL SCREW, 30mm 4 Valor Medical Solutions South Texas (TX) CE118 506-04-122 5035A1018 ALTIVATE REVERSE TORX PERIPHERAL SCREW, 22mm 6 506-04-126 5037A1017 ALTIVATE REVERSE TORX PERIPHERAL SCREW, 26mm 17 Matrix Orthopedics Surgical Devices Dallas/Fort Worth, TX CE120 506-04-122 5035A1018 ALTIVATE REVERSE TORX PERIPHERAL SCREW, 22mm 1 Joint Ventures Medical LLC Wisconsin (WI) CE121 506-04-122 5035A1018 ALTIVATE REVERSE TORX PERIPHERAL SCREW, 22mm 24 506-04-126 5037A1017 ALTIVATE REVERSE TORX PERIPHERAL SCREW, 26mm 1 WHISKEY TRAIL ORTHOPEDICS LLC - AUS Tennessee (TN) CE123 506-04-122 5035A1018 ALTIVATE REVERSE TORX PERIPHERAL SCREW, 22mm 8 506-04-126 5037A1017 ALTIVATE REVERSE TORX PERIPHERAL SCREW, 26mm 4 McCants Medical, LLC Florida (FL) CE128 506-04-122 5035A1018 ALTIVATE REVERSE TORX PERIPHERAL SCREW, 22mm 4 506-04-126 5037A1017 ALTIVATE REVERSE TORX PERIPHERAL SCREW, 26mm 2 Florida Direct - FSW Medical Florida (FL) FL001 506-04-118 5033A1020 ALTIVATE REVERSE TORX PERIPHERAL SCREW, 18mm 16 506-04-122 5035A1018 ALTIVATE REVERSE TORX PERIPHERAL SCREW, 22mm 14 506-04-126 5037A1011 ALTIVATE REVERSE TORX PERIPHERAL SCREW, 26mm 6 506-04-126 5037A1017 ALTIVATE REVERSE TORX PERIPHERAL SCREW, 26mm 72 506-04-130 5039A1015 ALTIVATE REVERSE TORX PERIPHERAL SCREW, 30mm 27 506-04-130 5039A1021 ALTIVATE REVERSE TORX PERIPHERAL SCREW, 30mm 65 Great Lakes Medical Illinois (IL) GL108 506-04-118 5033A1020 ALTIVATE REVERSE TORX PERIPHERAL SCREW, 18mm 8 506-04-122 5035A1018 ALTIVATE REVERSE TORX PERIPHERAL SCREW, 22mm 6 506-04-126 5037A1011 ALTIVATE REVERSE TORX PERIPHERAL SCREW, 26mm 2 506-04-126 5037A1017 ALTIVATE REVERSE TORX PERIPHERAL SCREW, 26mm 14 Certus Medical LLC Indiana (IN) & Michigan (MI) GL112 506-04-122 5035A1018 ALTIVATE REVERSE TORX PERIPHERAL SCREW, 22mm 14 506-04-126 5037A1011 ALTIVATE REVERSE TORX PERIPHERAL SCREW, 26mm 17 506-04-126 5037A1017 ALTIVATE REVERSE TORX PERIPHERAL SCREW, 26mm 3 506-04-130 5039A1021 ALTIVATE REVERSE TORX PERIPHERAL SCREW, 30mm 23 I & S Medical West Virginia (WV) & Kentucky (KY) GL113 506-04-126 5037A1011 ALTIVATE REVERSE TORX PERIPHERAL SCREW, 26mm 1 506-04-130 5039A1015 ALTIVATE REVERSE TORX PERIPHERAL SCREW, 30mm 8 Osteo Solutions Ohio (OH), Kentucky (KY), GL114 506-04-122 5035A1018 ALTIVATE REVERSE TORX PERIPHERAL SCREW, 22mm 15 506-04-126 5037A1011 ALTIVATE REVERSE TORX PERIPHERAL SCREW, 26mm 16 506-04-126 5037A1017 ALTIVATE REVERSE TORX PERIPHERAL SCREW, 26mm 6 506-04-130 5039A1015 ALTIVATE REVERSE TORX PERIPHERAL SCREW, 30mm 32 Procore Medical, LLC North Carolina (NC) MA102 506-04-118 5033A1020 ALTIVATE REVERSE TORX PERIPHERAL SCREW, 18mm 1 506-04-122 5035A1018 ALTIVATE REVERSE TORX PERIPHERAL SCREW, 22mm 14 506-04-126 5037A1017 ALTIVATE REVERSE TORX PERIPHERAL SCREW, 26mm 2 506-04-130 5039A1015 ALTIVATE REVERSE TORX PERIPHERAL SCREW, 30mm 59 Medsource Rhode Island (RI), Massachusetts (MA) NE131 506-04-122 5035A1018 ALTIVATE REVERSE TORX PERIPHERAL SCREW, 22mm 3 506-04-126 5037A1011 ALTIVATE REVERSE TORX PERIPHERAL SCREW, 26mm 12 506-04-126 5037A1017 ALTIVATE REVERSE TORX PER
Product codes and lots
Lot Code: Item: 506-04-126 GTIN: 00190446829850 Lot\s: 5037A1011 / 5037A1017

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2008-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2008-2025Class IIOngoingVoluntary: Firm initiated
Z-2009-2025Class IIOngoingVoluntary: Firm initiated
Z-2010-2025Class IIOngoingVoluntary: Firm initiated
Z-2011-2025Class IIOngoingVoluntary: Firm initiated