Z-1914-2025Class II
Long Nail, RIGHT Gamma4 10x320mm x 125, component of the Gamma4 System. Catalog Number: 84250320S
Stryker Corp. / initiated 2025-05-28 / Open, Classified / root cause: manufacturing process
Stryker GmbH began a recall of Long Nail, RIGHT Gamma4 10x320mm x 125, component of the Gamma4 System. Catalog Number: 84250320S on . FDA classified it as Class II and gives the reason as "Investigation confirmed that the GAMMA4 Right Long Nail was manufactured with the geometry of a left nail". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1914-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Stryker Corp.
- Product
- Long Nail, RIGHT Gamma4 10x320mm x 125, component of the Gamma4 System. Catalog Number: 84250320S.
- Product code
- HSB Rod, Fixation, Intramedullary And Accessories
- Reason for recall
- Investigation confirmed that the GAMMA4 Right Long Nail was manufactured with the geometry of a left nail.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- Stryker notified consignees on about 05/28/2025 via FedEx letter. Consignees were instructed to review inventory and immediately quarantine any affected units on hand, and complete and return the Business Reply Form documenting receipt and product segregation. Upon receipt of the response form, Stryker will contact consignees to arrange for return of the affected units. If affected products have been further distributed, consignees were instructed to notify those customers and let Stryker know.
- Distribution
- Worldwide - US Nationwide distribution in the states of AL, FL, NJ, OR and the countries of Australia and Japan.
- Product codes and lots
- Catalog Number: 84250320S. GTIN: 07613327479515. Lot Number: K1BC9B7.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|