Z-2085-2025Class I

Brand Name: FEM-FLEXII Product Name: Femoral Arterial Cannula Model/Catalog Number: FEMII018AS Software...

Edwards Lifesciences LLC / initiated 2025-05-14 / Open, Classified

Edwards Lifesciences, LLC began a recall of Brand Name: FEM-FLEXII Product Name: Femoral Arterial Cannula Model/Catalog Number: FEMII018AS Software Version: N/A Product Description: The Edwards FEMFLEXII... on . FDA classified it as Class I and gives the reason as "due to complaints received for 3mm to 4mm section wire exposed/protruding from the outside of the cannula body". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-2085-2025
Classification
Class I
Status
Open, Classified
Initiated
Posted
Recalling firm
Edwards Lifesciences LLC
Product
Brand Name: FEM-FLEXII Product Name: Femoral Arterial Cannula Model/Catalog Number: FEMII018AS Software Version: N/A Product Description: The Edwards FEMFLEXII Arterial Perfusion Cannluae provides or establishes arterial perfusion in the extracorporeal circuit for d 6 hours. Component: No
Product code
DQR Cannula, Catheter
Reason for recall
due to complaints received for 3mm to 4mm section wire exposed/protruding from the outside of the cannula body.
FDA root cause
Under Investigation by firm
Action
On 05/16/2025, and 05/19/2025, the firm sent via FedEx overnight an "URGENT: PRODUCT RECALL" Letter to customers informing them that Edwards has confirmed a small number of occurrences impacting the OptiSite Arterial Perfusion cannula in which a 3mm to 4mm section of wire, from the wire-reinforcement coil at the cannula tip, was found to be released from the cannula body. The risk for the wire-reinforcement coil protruding out of the cannula body is major tissue damage and hemolysis. Customers are instructed to: " Verify their inventory on the attached customer acknowledgement form. " Share this notice with appropriate clinical staff at their site. " No patient follow-up or notification is necessary. " Return a completed Customer Acknowledgment Form to Edwards eCV Representative or via email to Edwards Customer Service at [email protected] within 15 days of receipt of this notification. " After receiving RGA number, return any impacted product to Edwards at the address below. A credit will be issued upon receipt of returned product. enableCV Distribution Center 6644 W. 2100 S. Suite D West Valley, UT 84128 For questions/assistance contact - Customer Service at 1-888-943-2783, or email: [email protected]
Quantity in commerce
558
Distribution
Worldwide - US Nationwide distribution including in the states of ALABAMA, ARIZONA, ARKANSAS, CALIFORNIA, COLORADO, CONNECTICUT, DELAWARE, DISTRICT OF COLUMBIA, FLORIDA, GEORGIA, HAWAII, IDAHO, ILLINOIS, INDIANA, IOWA, KANSAS, KENTUCKY, LOUISIANA, MAINE, MARYLAND, MASSACHUSETTS, MICHIGAN, MINNESOTA, MISSISSIPPI, MISSOURI, MONTANA, NEBRASKA, NEVADA, NEW HAMPSHIRE, NEW JERSEY, NEW MEXICO, NEW YORK, NORTH CAROLINA, NORTH DAKOTA, OHIO, OKLAHOMA, OREGON, PENNSYLVANIA, Puerto Rico, RHODE ISLAND, SOUTH CAROLINA, SOUTH DAKOTA, TENNESSEE, TEXAS, UTAH, VIRGINIA, WASHINGTON, WEST VIRGINIA, WISCONSIN and the countries of Albania, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, CANARY ISLANDS, Chile, Colombia, Cyprus, Czech Republic, Denmark, France, Germany, Greece, Hungary, Iceland, India, Ireland, Israel, Italy, Japan, Luxembourg, Malaysia, Malta, Netherlands, Northern Ireland, Norway, Pakistan, Panama, Poland, Portugal, Romania, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, United Arab Emirates (UAE), United Kingdom.
Product codes and lots
Lot Code: Model No: FEMII018AS; UDI: 00690103168358; All lots up to: BSLC9064 Serial Numbers: 131776 BSLC4190 BSLC9064

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2085-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2080-2025Class IOngoingVoluntary: Firm initiated
Z-2081-2025Class IOngoingVoluntary: Firm initiated
Z-2082-2025Class IOngoingVoluntary: Firm initiated
Z-2083-2025Class IOngoingVoluntary: Firm initiated
Z-2084-2025Class IOngoingVoluntary: Firm initiated
Z-2085-2025Class IOngoingVoluntary: Firm initiated