Z-2497-2025Class II

LIFEPAK 15 Part / Catalog Numbers / UDI-DI code: V15-7-000022 / 99577-001957 / 883873871690 V15-2-000051 /...

Physio Control, Inc. / initiated 2025-07-22 / Open, Classified

Physio-Control, Inc. began a recall of LIFEPAK 15 Part / Catalog Numbers / UDI-DI code: V15-7-000022 / 99577-001957 / 883873871690 V15-2-000051 / 99577-000046 / 883873988022 V15-2-000099 /... on . FDA classified it as Class II and gives the reason as "Due to required inspections not being performed on products/units that have gone through servicing". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-2497-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Physio Control, Inc.
Product
LIFEPAK 15 Part / Catalog Numbers / UDI-DI code: V15-7-000022 / 99577-001957 / 883873871690 V15-2-000051 / 99577-000046 / 883873988022 V15-2-000099 / 99577-000093 / 883873979020 V15-2-001603 / 99577-001256 / 883873911631 LIFEPAK 15: The LP15 is intended for use by trained medical personnel in out-of-doors and indoor emergency care settings within the environmental conditions specified in the Operating Instructions. The LP15 is designed to be used during ground transportation except when specified otherwise. Manual mode monitoring and therapy functions are intended for use on adult and pediatric patients. Automated external defibrillation mode is intended for use on patients eight years of age and older.
Product code
MKJ Automated External Defibrillators (Non-Wearable)
Reason for recall
Due to required inspections not being performed on products/units that have gone through servicing.
FDA root cause
Under Investigation by firm
Action
On 07/22/2025, the firm sent via certified mail an "URGENT: Medical device recall" Letter to customers informing them that, Stryker discovered a subset of work orders performed between August 10th 2022 and October 30th 2023 with Performance Inspection Procedures (PIP) recorded as PASS without the corresponding objective evidence properly documented in the work order to ensure the devices are performing as intended. Customers are instructed to: 1. Immediately check their inventory to locate the affected product as listed in the letter. 2. On the business reply form, customer need to indicate if their devices have been serviced since October 30th 2023 (after that time, if affected device have undergone Preventative Maintenance through Stryker, Stryker has verified that the PIPs were completed and passed in accordance with the required standards. For these devices No additional action is required. 3. Return business reply form. Upon receipt, Stryker will contact customers to arrange for corrective action on the device if service has not been performed. 4. There are no additional actions that users can and should take. Continue to follow the Operating Instructions and perform the daily checks and user tests. Tests can be found in the chapter "Maintaining the Equipment' subsection "General Maintenance and Testing" For Questions or concerns, contact Customer Service at +1 800 787-9537, option 2, from 8:00 AM to 7:00 PM (Eastern Time, Monday thru Friday or email [email protected]
Quantity in commerce
4 units
Distribution
U.S. Nationwide distribution in the states of CO and MT.
Product codes and lots
Part / Catalog Numbers / UDI-DI code / Serial Number: V15-7-000022 / 99577-001957 / 883873871690 / 49085553 V15-2-000051 / 99577-000046 / 883873988022 / 38154524 V15-2-000099 / 99577-000093 / 883873979020 / 39059475 V15-2-001603 / 99577-001256 / 883873911631 / 40973093

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2497-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2497-2025Class IIOngoingVoluntary: Firm initiated
Z-2498-2025Class IIOngoingVoluntary: Firm initiated
Z-2499-2025Class IIOngoingVoluntary: Firm initiated