Endosee System Convenience Kit. Model Number: ES-TRAY. Endosee System Convenience Kit
CooperSurgical, Inc. began a recall of Endosee System Convenience Kit. Model Number: ES-TRAY. Endosee System Convenience Kit on . FDA classified it as Class II and gives the reason as "The HSG Procedure Trays and Endosee System Convenience Kits contain non-sterile iodine pouches, however the peel away lid labels incorrectly identify the iodine pouches as sterile". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2136-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- CooperSurgical, Inc.
- Product
- Endosee System Convenience Kit. Model Number: ES-TRAY. Endosee System Convenience Kit
- Product code
- HIB Speculum, Vaginal, Nonmetal
- Reason for recall
- The HSG Procedure Trays and Endosee System Convenience Kits contain non-sterile iodine pouches, however the peel away lid labels incorrectly identify the iodine pouches as sterile.
- Root cause tag
- labeling and UDI
- FDA root cause
- Under Investigation by firm
- Action
- CooperSurgical notified consignees on about 06/11/2025 via letter sent through certified mail. Specifically, consignees that are physicians or medical facilities were instructed to identify and quarantine any affected product in inventory, complete and return the provided response form, and work with CooperSurgical to arrange for the return of affected units. Distributors were instructed to identify and quarantine any affected product in inventory, complete and return the provided response form, and work with CooperSurgical to arrange for the return of affected units. They were also instructed to notify customers if the product was further distributed.
- Quantity in commerce
- 991 units
- Distribution
- US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and Washington DC.
- Product codes and lots
- Model Number: ES-TRAY. UDI-DI: 60888937027268. Lot Numbers: 335995, 339340, 339341, 339342, 339343, 339344, 339345, 343008.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2135-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2136-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2137-2025 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.