Z-2133-2025Class II
Signa Pet/mr
GE Healthcare / initiated 2025-05-23 / Open, Classified / root cause: design
GE Medical Systems, LLC began a recall of SIGNA PET/MR on . FDA classified it as Class II and gives the reason as "GE HealthCare has become aware that gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning. As a result, acoustic...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2133-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- GE Healthcare
- Product
- Signa Pet/mr
- Product code
- LNH System, Nuclear Magnetic Resonance Imaging
- Reason for recall
- GE HealthCare has become aware that gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning. As a result, acoustic levels can exceed 99dB, the limit established by International Electrotechnical Commission (IEC 60601-2-33) when using hearing protection with a Noise Reduction Rating (NRR) of 29dB, as currently required in the Operator Manual.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION to its consignees on 05/23/2025 delivered using traceable means. The notice explained the issue, risk, and requested the following: Actions to be taken by Customer/User: You can continue to use your MR system by following the instructions below: Ensure hearing protection with a Noise Reduction Rating (NRR) of at least 33dB is used during exams on these systems. 200 pairs are provided with this letter. Please post this letter in your facility on or near the MR operator console. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please retain this document for your records. Please complete and return the attached acknowledgement form to [email protected].
- Distribution
- Worldwide distribution - US Nationwide and the country of China.
- Product codes and lots
- GTIN: 00840682105378
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2130-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2131-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2132-2025 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2133-2025 | Class II | Ongoing | Voluntary: Firm initiated |