Z-2073-2025Class II

VITROS Chemistry Products OP Reagent Gen 15. Model Number: 680 1997. The VITROS OP Reagent is a...

Ortho-Clinical Diagnostics, Inc. / initiated 2025-06-12 / Open, Classified

Ortho-Clinical Diagnostics, Inc. began a recall of VITROS Chemistry Products OP Reagent Gen 15. Model Number: 680 1997. The VITROS OP Reagent is a dual-chambered package containing ready-to-use liquid reagents... on . FDA classified it as Class II and gives the reason as "Lots from Generation (GEN) 15 of VITROS Chemistry Products OP Reagent may generate lower than expected quality control (QC) results. If the customer is unable to get passing quality control results, then the customer...". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-2073-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
VITROS Chemistry Products OP Reagent Gen 15. Model Number: 680 1997. The VITROS OP Reagent is a dual-chambered package containing ready-to-use liquid reagents that are used to detect opiates in urine. One sales unit contains 6 Microtip packs of reagent. Each pack has 50 OP tests.
Product code
DJG Enzyme Immunoassay, Opiates
Reason for recall
Lots from Generation (GEN) 15 of VITROS Chemistry Products OP Reagent may generate lower than expected quality control (QC) results. If the customer is unable to get passing quality control results, then the customer would be unable to run the OP-LO protocol leading to a potential delay in patient results.
FDA root cause
Under Investigation by firm
Action
Quidel Ortho notified consignees on about 06/12/2025 via FedEx letter. Distributors were instructed to discontinue distribution and discard inventory of any affected lots, send the provided customer letter and Confirmation of Receipt to any affected customers, and complete and return their Confirmation of Receipt. Customers were instructed that if their laboratory is unable to generate QC results within acceptable range using lots from GEN 15 of VITROS OP Reagent, discontinue using, render unusable, and discard your inventory of VITROS OP Reagent, GEN 15. Additionally, there were requested to complete and return the Confirmation of Receipt, maintain the notification with all records, and provide the notification if affected units were further distributed.
Quantity in commerce
2,017 units
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Denmark, France, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore, Spain, Sweden, The Netherlands, United Kingdom.
Product codes and lots
Model Number: 680 1997. UDI-DI: 10758750001835. Lot Numbers: 1527-15-2749 (exp. 17-Nov-2025), 1527-15-2771 (exp. 17-Dec-2025), 1527-15-2818 (exp. 11-Apr-2026), 1527-15-2871 (exp. 15-May-2026).

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2073-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2073-2025Class IIOngoingVoluntary: Firm initiated