TrueDent Yellow, TDM140, Model No. OBJ-09169 Light-curable methacrylate-based resin that enables fabrication...
STRATASYS LTD began a recall of TrueDent Yellow, TDM140, Model No. OBJ-09169 Light-curable methacrylate-based resin that enables fabrication of dental appliances on . FDA classified it as Class III and gives the reason as "Customers unable to use cartridges due to formatting error in expiration date which leads to switching the day with the month. Or, the system will not recognize that a resin is expired and will not prevent the user from...". As of , the recall status on record is Completed, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2093-2025
- Classification
- Class III
- Status
- Completed
- Initiated
- Posted
- Recalling firm
- Stratasys Inc.
- Product
- TrueDent Yellow, TDM140, Model No. OBJ-09169 Light-curable methacrylate-based resin that enables fabrication of dental appliances
- Product code
- EBI Resin, Denture, Relining, Repairing, Rebasing
- Reason for recall
- Customers unable to use cartridges due to formatting error in expiration date which leads to switching the day with the month. Or, the system will not recognize that a resin is expired and will not prevent the user from using an expired material.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The firm initiated the recall to the consignee level on January 15, 2024. Customers were asked to coordinate a cartridge replacement. The reporting of this recall was done by the firm retroactively; the issue was resolved in January 2024.
- Quantity in commerce
- 68
- Distribution
- US Nationwide distribution including in the states of Arizona, Arkansas, California, Colorado, Connecticut, Florida, Georgia, Illinois, Indiana, Maryland, Massachusetts, Michigan, Nebraska, New York, North Carolina, North Dakota, Oklahoma, Oregon, Utah, Wisconsin.
- Product codes and lots
- UDI-DI 7290018915806 Lot 22861 (exp. 01-Feb-2024)
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2090-2025 | Class III | Completed | Voluntary: Firm initiated | ||
| Z-2091-2025 | Class III | Completed | Voluntary: Firm initiated | ||
| Z-2092-2025 | Class III | Completed | Voluntary: Firm initiated | ||
| Z-2093-2025 | Class III | Completed | Voluntary: Firm initiated | ||
| Z-2094-2025 | Class III | Completed | Voluntary: Firm initiated |