Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2462-2025Class II

Bardex Lubri-Sil I.C. 400-series Temperature Sensing Silver/Hydrogel Coated All-Silicone Foley Catheter 5cc...

C.R. Bard Inc / initiated 2025-06-18 / Open, Classified / root cause: manufacturing process

C.R. Bard Inc began a recall of Bardex Lubri-Sil I.C. 400-series Temperature Sensing Silver/Hydrogel Coated All-Silicone Foley Catheter 5cc Balloon, REF: 119314; SureStep Foley Tray System... on . FDA classified it as Class II and gives the reason as "Foley catheter may have an obstructed lumen, which may result in an inability to drain the bladder, of urine and may include pelvic or abdominal pain, flank pain, bladder dysfunction, obstructive uropathy, acute kidney...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2462-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
C.R. Bard Inc
Product
Bardex Lubri-Sil I.C. 400-series Temperature Sensing Silver/Hydrogel Coated All-Silicone Foley Catheter 5cc Balloon, REF: 119314; SureStep Foley Tray System Lubri-Sil I.C. Complete Care, REF: A319514A; Bard Silver Lubri-Sil Foley Tray Temperature Sensing with Bard Hydrogel and Bacti-Guard Silver Alloy Coating, 14Fr, REF: 29000J14, 29030J14, 29000J14S; Bard Silver Lubri-Sil Foley Tray Temperature Sensing, Complete Care with Bard Hydrogel and Bacti-Guard Silver Alloy Coating, 14Fr, REF: 29030J14S
Product code
EZL Catheter, Retention Type, Balloon
Reason for recall
Foley catheter may have an obstructed lumen, which may result in an inability to drain the bladder, of urine and may include pelvic or abdominal pain, flank pain, bladder dysfunction, obstructive uropathy, acute kidney injury and/or hydronephrosis.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
On 6/18/2025, recall notices were mailed or emailed to Recall Coordinators, Directors of Nursing, Purchasing, Risk Management, and distributors who were informed the following: The affected product was distributed from 07 March 2024 to 14 May 2025. - If affected product is currently indwelling and draining urine appropriately, the catheter does not need to be removed. - For affected product currently not in use, quarantine, and dispose of unused product and source alternative product. - Share this notice with any users of the product within your facilities to ensure they are also aware of this Medical Device Recall (Correction). - If you have purchased this product through a distributor, please contact this distributor to request appropriate credit or order alternate replacement product. - Complete and return the attached Customer Response Form via Email [email protected] Report any complaints experienced with the use of this product to the firm via the North American Regional Complaint Center: Phone: 1-844-8BD-LIFE (1-844-823-5433) say "Product Complaints" when prompted Mon-Fri 8:00am and 5:00pm CT, Email: [email protected] Firm's Medical Information Services: 1-800-555-7422, [email protected]
Quantity in commerce
175,158
Distribution
Worldwide - US Nationwide distribution in the states of IA, PA, AR, OR, MN, IN, WV, OK, AL, MD, HI, GA, AZ, MA, FL, CA and the countries of Japan.
Product codes and lots
Catalog/UDI-DI/Lot(Expiration): 119314/00801741039843, 10801741039840/NGJS3588(31/03/2029), NGKQ2500(21/12/2029), NGKQ0170(30/11/2029), NGJW2499(31/08/2029), NGJT1284 (31/03/2029), NGKQ2492(31/12/2029), NGKP3152(31/12/2029), NGJX0215(31/07/2029). A319514A/00801741100321, 10801741100328/NGJN0498(31/01/2026), NGJW3127(30/04/2027) 29000J14/21200BZY00110000/MYJX0683(28/05/2027), MYJP0751(28/02/2027), MYJR1984(28/04/2027), MYJQ0070(28/04/2027), MYJS6278(28/04/2027), MYJR3026(28/04/2027), MYJT3766(28/04/2027), MYJY6197(28/06/2027), MYJX2833(28/06/2027), MYJQ4022(28/04/2027), MYJT3764(28/04/2027), MYJZ0962(28/06/2027), MYJS4679(28/04/2027), MYJX0686(28/06/2027), MYJV4359(28/03/2027), MYJS3597(28/04/2027), MYJW1900(28/04/2027), MYJT3765(28/04/2027), MYJY1245(28/06/2027), MYJY1242(28/06/2027), MYJY1248(28/06/2027). 29030J14/21200BZY00110000/MYJX2846(28/06/2027), MYJP2013(28/02/2027), MYJT1017(28/04/2027), MYJW1906(28/04/2027), MYJT4535(28/07/2027), MYJP4407(28/04/2027), MYJS1895(28/04/2027), MYJW1904(28/04/2027), MYJV4102(28/03/2027), MYJX2836(28/05/2027), MYJZ4526(28/06/2027), MYJT3468(28/04/2027), MYJP2012(28/02/2027), MYJV4099(28/03/2027), MYJZ0964(28/06/2027), MYJQ0071(28/04/2027), MYJS6279(28/04/2027), MYJY4511(28/06/2027), MYJQ4024(28/04/2027), MYJZ0963(28/06/2027), MYJT1018(28/04/2027), MYJR3027(28/04/2027), MYJS3599(28/04/2027), MYJQ4023(28/04/2027), MYJR3083(28/04/2027), MYJY4512(28/06/2027), MYJQ0073(28/04/2027), MYJS1893(28/04/2027), MYJN2580(28/01/2027), MYJU2705(28/08/2027), MYJZ4527(28/06/2027), MYJS1894(28/04/2027), MYJV4103(28/03/2027), MYJR3059(28/04/2027), MYJU0772(28/08/2027), MYJV4101(28/03/2027), MYJY4513(28/06/2027), MYJQ0072(28/04/2027), MYJU3846(28/03/2027), MYJU3847(28/05/2027), MYJX2841(28/06/2027), MYJX2840(28/05/2027), MYJX2839(28/05/2027), MYJY4510(28/06/2027), MYJY4509(28/06/2027), MYJV4100(28/06/2027), MYJU3848(28/06/2027). 29000J14S/21200BZY00110000/MYJY1249(28/06/2027), MYJY0937(28/05/2027), MYJW1903(28/03/2027). 29030J14S/21200BZY00110000/MYJU0773(28/08/2027)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2462-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2462-2025Class IIOngoingVoluntary: Firm initiated