Z-2134-2025Class II
Azurion 7 M20. Fluoroscopic X-Ray System
Philips Medical Systems Nederland B.V. / initiated 2025-06-04 / Open, Classified / root cause: design
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. began a recall of Azurion 7 M20. Fluoroscopic X-Ray System on . FDA classified it as Class II and gives the reason as "Potential for bolts connecting the gearbox to the mounting flange of the C-arm to become loose". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2134-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Philips Medical Systems Nederland B.V.
- Product
- Azurion 7 M20. Fluoroscopic X-Ray System.
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- Potential for bolts connecting the gearbox to the mounting flange of the C-arm to become loose.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- A "Device URGENT Medical Device Correction" notification dated 6/03/2025 was mailed to consignees. Consignees are instructed to circulate the provided notification to all system users and keep a copy of the recall notification with affected devices to maintain awareness. If you notice unusual noise during C-arm roll motion, or if the C-arm roll motion is disabled and the user message Motorized movement is not available appears, immediately contact a Philips representative. Keep this Urgent Medical Device Correction letter with the documentation of the system until Philips corrects your system. Ensure that the letter is in a place likely to be seen/viewed. Affected systems may continue to be used in accordance with their Instructions for Use (IFU). Complete and return the response form included in this Urgent Medical Device letter. Consignees are provided instructions to avoid rapidly pressing and releasing footswitch pedals and how to restart systems that experience the failure mode. The firm will deploy a software update in all affected system by June 2025. Consignees with any questions should contact their local Philips representative at 1-800-722-9377.
- Quantity in commerce
- 78 units
- Distribution
- Worldwide distribution: US (nationwide) to states of: CA, CO, DC, DE, FL, IL, IN, KY, MA, MD, MN, MO, MT, NJ, NV, NY, OH, SC, TX, WA; and OUS (Foreign) to countries of: Argentina (AR), Australia (AU), Canada (CA), Chile (CL), Germany (DE), Dominican Republic (DO), Egypt (EG), ES, France (FR), United Kingdom (GB), Italy (IT), Japan (JP), Korea, Republic (KR), Mexico (MX), Netherlands (NL), New Zealand (NZ), Oman (OM), Reunion (RE), Saudi Arabi (SA), Thailand (TH), Viet Nam (VN).
- Product codes and lots
- System Model No. 722079; UDI: 00884838085268; Serial No. 525, 584, 587, 613, 631, 658, 708, 774, 777, 793, 800, 801, 838, 842, 843, 862, 865, 866, 879, 881, 896, 899, 900, 905, 908, 909, 912, 915, 923, 925, 930, 939, 948, 955, 957, 992, 750, 814, 321, 370, 549, 567, 670, 687, 688, 693, 707, 743, 759, 767, 768, 775, 785, 786, 791, 819, 855, 871, 872, 902, 916, 917, 933, 936, 956, 960, 963, 965, 966, 969, 977, 993.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2134-2025 | Class II | Ongoing | Voluntary: Firm initiated |