Z-2179-2025Class I

BD Alaris Pump Module Model, 8100 with Guardrails, Suite MX software versions

CareFusion 303, Inc / initiated 2025-07-08 / Open, Classified

CareFusion 303, Inc. began a recall of BD Alaris Pump Module Model, 8100 with Guardrails, Suite MX software versions on . FDA classified it as Class I and gives the reason as "Infusion pump module used with compatible pump infusion sets may perform outside the established performance ranges for flow rate and bolus accuracy, downstream and upstream occlusion time to alarm, and post-occlusion...". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-2179-2025
Classification
Class I
Status
Open, Classified
Initiated
Posted
Recalling firm
CareFusion 303, Inc
Product
BD Alaris Pump Module Model, 8100 with Guardrails, Suite MX software versions
Product code
FRN Pump, Infusion
Reason for recall
Infusion pump module used with compatible pump infusion sets may perform outside the established performance ranges for flow rate and bolus accuracy, downstream and upstream occlusion time to alarm, and post-occlusion bolos volume.
FDA root cause
Under Investigation by firm
Action
On December 12, 2025, CareFusion 303, Inc. (BD) issued a updated "Urgent Medical device Recall Notice via E-Mail and physical mail. The updated notice includes important information on infusion sets as well as labeling updates. Previous notifications were sent on July 8,2025, July 17, 2025, September 11, 2025 and September 12, 2025. CareFusion ask consignees to take the following actions: 1. Review your inventory for the catalog numbers included in the updated letter. Remove and discard any remaining inventory per your facility guidelines. 2.To receive credit or replacement product, please return the completed customer response form attached to this notice. 3.Please contact your BD Clinical Consultants and Account Executives for alternative sets information. 4. Review the updated sets performance data included in the letter. 5. Circulate this notice within your facility network to ensure that all concerned personnel are made aware of this issue. 6. Distributors should provide a copy of this notice to all customers who may have purchased an Alaris Pump Module. If you require further assistance, please contact: BD Technical Support at Phone: 888-812-3229; Phone hours: 6:00am to 5:00pm PT Monday Friday and Email: [email protected].
Quantity in commerce
33,162,680 units total
Distribution
US Nationwide Distribution to states of: AK AL AR AS AZ CA CO CT DC DE FL GA GU HI IA ID IL IN KS KY LA MA MD ME MI MN MO MP MS MT NC ND NE NH NJ NM NV NY OH OK OR PA PR RI SC SD TN TX UT VA VT WA WI WV WY; and OUS (foreign) to countries of: Japan, Italy, Germany, Turkey
Product codes and lots
BD Alaris" Pump Module Model 8100- All Serial Numbers/ UDI(s): 10885403222054, 10885403517723, 10885403810015, 10885403810039, 10885403810046

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2179-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2179-2025Class IOngoingVoluntary: Firm initiated
Z-2180-2025Class IOngoingVoluntary: Firm initiated