Z-2177-2025Class II

Dual Flat Panel Monitor (DFPM) accessories for HarmonyAIR A-Series Surgical Lighting Systems

STERIS Corporation / initiated 2025-06-23 / Open, Classified / root cause: manufacturing process

Steris Corporation began a recall of Dual Flat Panel Monitor (DFPM) accessories for HarmonyAIR A-Series Surgical Lighting Systems on . FDA classified it as Class II and gives the reason as "The DFPM yoke assembly ("yoke assembly") may detach from the spring arm. Should the yoke assembly detach from the spring arm, the yoke assembly would not be expected to fully detach but rather hang from the wiring...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2177-2025
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
STERIS Corporation
Product
Dual Flat Panel Monitor (DFPM) accessories for HarmonyAIR A-Series Surgical Lighting Systems
Product code
FSY Light, Surgical, Ceiling Mounted
Reason for recall
The DFPM yoke assembly ("yoke assembly") may detach from the spring arm. Should the yoke assembly detach from the spring arm, the yoke assembly would not be expected to fully detach but rather hang from the wiring within the yoke assembly.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Steris issued an URGENT MEDICAL DEVICE RECALL NOTICE dated 06/17/2025 to its consignees on 06/23/2025 disseminated via FedEx. The notice explained the issue, potential risk, and requested the following: User Action: Please ensure the following steps are completed: 1. Review the revised Preventive Maintenance Manual and Interval Based Check List language depending on which HarmonyAIR Surgical Lighting System model(s) are in use at your facility (A-Series and/or E-Series) (please reference Attachments B-E of the letter for maintenance instructions). 2. Immediately complete the maintenance activities as instructed, and then again annually going forward. 3. Complete the Medical Device Recall Response Form included with this letter. 4. Return the completed Response Form via email to: [email protected], or via fax to 440-392-8963. 5. Distributors: If you have further distributed this product, please identify your Customer(s), and notify them of this recall.
Quantity in commerce
167 units
Distribution
Worldwide - US Nationwide and Puerto Rico and the country of Canada.
Product codes and lots
UDI: 00724995203245: Serial Numbers: 0400224028, 0401625167, 0401625171, 0401625180, 0401625185, 0401625190, 0402022165, 0402022169, 0402422003, 0402622007, 0402925148, 0403122053, 0403122057, 0403125009, 0403223079, 0403223083, 0403223087, 0403223091, 0403321066,0403321070, 0403423063, 0403423067, 0403423071, 0403423075, 0403423079, 0403423083, 0403423087, 0404022003, 0404622011, 0404622014, 0404622017, 0404622020, 0404622023, 0404822104, 0405025032, 0405924100, 0405924103, 0406624134, 0406624138, 0406624142, 0406722052, 0406722056, 0406722060, 0406722064, 0406823005, 0407625022, 0408025008, 0408222120, 0408222124, 0408222128, 0408222132, 0408922093, 0408922097, 0408922101, 0409622008, 0409822014, 0409822018, 0410822042, 0410822046, 0410823068, 0410923149, 0410923152, 0410923155, 0413223057, 0413223060, 0414223170, 0414223174, 0414223178, 0414223182, 0414223186, 0414223190, 0414223194, 0414223198, 0414323003, 0414323006, 0415223065, 0415223068, 0415223071, 0415223074, 0415923036, 0416924069, 0416924074, 0417422053, 0418023091, 0418023095, 0418023099, 0418023103, 0418023107, 0418023111, 0419223082, 0419523010, 0419724177, 0419924007, 0420223048, 0420223051, 0420223054, 0420223057, 0420223060, 0420823170, 0421322041, 0421322045, 0421623206, 0421623211, 0421623216, 0421623221, 0421623226, 0421623231, 0421623236, 0421623241, 0422923010, 0423023059, 0423023304, 0423023307, 0423023379, 0423023384, 0423023389, 0423023394, 0423023399, 0423023451, 0423422042, 0423422045, 0423422048, 0423924004, 0424224127, 0424224131, 0424323096, 0426223004, 0426223008, 0427922038, 0427922042, 0428824102, 0428924075, 0428924079, 0428924083, 0429823019, 0429823023, 0430323199, 0430523194, 0430523197, 0430722034, 0431123070, 0432324031, 0432324156, 0432324159, 0432324162, 0432324165, 0432722064, 0433724005, 0433724010, 0433724015, 0434023084, 0434023087, 0434023090, 0434023111, 0434023116, 0434023121, 0434622027, 0434622031, 0434622035, 0434722013, 0434722159, 0434722163, 0436223068, 0436624179, 0436624184, 0436624189, 0436624194;

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2177-2025
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2177-2025Class IIOngoingVoluntary: Firm initiated
Z-2178-2025Class IIOngoingVoluntary: Firm initiated