Pentax Medical Video Processor; Model Number: EPK-i8020c
Pentax of America Inc began a recall of Pentax Medical Video Processor; Model Number: EPK-i8020c on . FDA classified it as Class II and gives the reason as "During endoscopic procedures using a combination of the video processor EPK-i8020c and i20c series video endoscope, the observed image can become reddish or dark. Users have observed smoke like steam and noted that the...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2329-2025
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- PENTAX of America, Inc.
- Product
- Pentax Medical Video Processor; Model Number: EPK-i8020c;
- Product code
- PEA Endoscope, Accessories, Image Post-Processing For Color Enhancement
- Reason for recall
- During endoscopic procedures using a combination of the video processor EPK-i8020c and i20c series video endoscope, the observed image can become reddish or dark. Users have observed smoke like steam and noted that the light guide at the tip is hot during use / after the endoscope is removed from the patient and may cause thermal injury to the patient s mucous membrane.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On 7/16/2025, "URGENT MEDICAL DEVICE CORRECTION" letters were sent to customers. Customer Instructions: PENTAX Medical will conduct the software update on the affected devices at your facility. Your local PENTAX Medical representative will contact you to schedule the required updates for your equipment. " Please download the Instructions for Use (IFU) for the EPK-i8020c Video Processor from the PENTAX Medical online IFU library at https://ifu.pentaxmedical.com. " Please complete the enclosed Field Correction Response Form upon receipt of this package, and email to PENTAX Medical at [email protected]. Contact Information: Please indicate through the attached response form that you have received and understood this information, by completing it and returning it no later than July18, 2025 at: [email protected] If you have any questions regarding this action, please feel free to contact us at: " Tel: 1-800-431-5880 (8:30 AM 5:00 PM, Monday Friday, EST) " Fax: (800)-579-5432) " Email: [email protected]
- Quantity in commerce
- 144 units
- Distribution
- US Distribution to states of: AZ, CA, DC, FL IA, ID, IL, KS, MA, MD, MI, MO, NC, NJ, NM, NY, OK, PA, TX, WA.
- Product codes and lots
- Model Number: EPK-i8020c; UDI-DI: 04961333247974; Serial Numbers: B0023Z1657, D0023Z0245, D0023Z0206, D0023Z0125, D0023Z0123, D0023Z0119, D0023Z0118, D0023Z0117, D0023Z0114, D0023Z0112, D0023Z0109, D0023Z0108, D0023Z0011, C0023Z0087, C0023Z0085, C0023Z0074, C0023Z0072, C0023Z0021, B0023Z1712, B0023Z1711, B0023Z1670, B0023Z1665, B0023Z1659, B0023Z1653, B0023Z1628, b0023z1501, B0023Z1195, B0023Z1707, C0023Z0084, B0023Z1732, B0023Z1669, D0023Z0137, D0023Z0141, D0158Z0049, D0158Z0033, E0023Z0022, D0158Z0053, D0158Z0019, E0023Z0026, E0023Z0008, D0023Z0248, D0023Z0241, D0023Z0222, D0023Z0155, B0023Z1276, C0023Z0082, D0023Z0233, B0023Z1461, D0023Z0139, D0023Z0140, D0023Z0122, B0023Z1709, E0023Z0020, E0023Z0009, E0023Z0002, D0023Z0142, E0023Z0024, E0023Z0018, E0023Z0013, E0023Z0004, C0023Z0095, C0023Z0078, E0023Z0007, E0158Z0057, E0158Z0035, B0023Z1735, B0023Z1717, B0023Z1648, B0023Z1639, B0023Z1631, C0158Z0047, C0158Z0042, B0023Z1481, B0023Z1478, B0023Z1451, B0023Z1335, B0023Z1193, B0023Z1288, B0023Z1258, B0023Z1296, B0023Z1290, B0023Z1286, B0023Z1280, B0023Z1279, B0023Z1256, B0023Z1250, B0023Z1245, B0023Z1676, B0023Z1573, B0023Z1565, B0023Z1656, B0023Z1694, D0158Z0055, B0023Z1577, B0023Z1570, B0023Z1682, B0023Z1678, B0023Z1661, B0023Z1637, D0158Z0048, D0158Z0047, E0023Z0016, E0023Z0001, D0023Z0071, D0023Z0150, D0023Z0148, D0023Z0143, B0023Z1277, D0023Z0152, D0023Z0144, D0023Z0106, D0023Z0102, B0023Z1667, B0023Z1652, B0023Z1645, B0023Z1727, B0023Z1723, B0023Z1702, E0023Z0003, D0023Z0244, D0023Z0232, C0023Z0091, D0023Z0208, B0023Z1655, D0023Z0219, D0023Z0017, D0023Z0008, C0023Z0098, C0023Z0097, C0023Z0096, C0023Z0094, C0023Z0092, B0023Z1716, B0023Z1698, B0023Z1697, B0023Z1696, B0023Z1671, B0023Z1255, B0023Z1247, B0023Z1291, B0023Z1289, B0023Z1287, B0023Z1285, B0023Z1243;
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2329-2025 | Class II | Ongoing | Voluntary: Firm initiated |